Selection of gamma analysis acceptance criteria in IMRT QA using Gafchromic EBT3 film dosimetry. Issue 2 (3rd December 2018)
- Record Type:
- Journal Article
- Title:
- Selection of gamma analysis acceptance criteria in IMRT QA using Gafchromic EBT3 film dosimetry. Issue 2 (3rd December 2018)
- Main Title:
- Selection of gamma analysis acceptance criteria in IMRT QA using Gafchromic EBT3 film dosimetry
- Authors:
- Isa Khan, Muhammad
Shakil, Muhammad
Tahir, Muhammad Bilal
Rafique, Muhammad
Iqbal, Tahir
Zahoor, Aliza
ur Rehman, Jalil
Iqbal, Khalid
Chow, James C. L. - Abstract:
- Abstract: Background and purpose: This study reported the justification and selection of acceptable γ criteria with respect to low (6 MV) and high (15 MV) photon beams for intensity-modulated radiation therapy quality assurance (IMRT QA) using the Gafchromic external beam therapy 3 (EBT3) film. Materials and methods: Five-field step-and-shoot IMRT was used to treat 16 brain IMRT patients using the dual-energy DHX-S linear accelerator (Varian Medical System, Palo Alto, CA, USA). Dose comparisons between computed values of the treatment planning system (TPS) and Gafchromic EBT3 film were evaluated based on γ analysis using the Film QA Pro software. The dose distribution was analysed with gamma area histograms (GAHs) generated using different γ criteria (3%/2 mm, 3%/3 mm and 5%/3 mm) for the 6 and 15 MV photon beams, to optimise the best distance-to-agreement (DTA) criteria with respect to the beam energy. Results: From the comparison between the dose distributions acquired from the TPS and EBT3 film, a DTA criterion of 3%/2 mm showed less dose differences (DDs) with passing rates up to 93% for the 6 MV photon beams, while for the 15 MV a relaxed DTA criterion of 5%/3 mm was consistent with the DD acceptability criteria with a 95% passing rate. Conclusions: Our results suggested that high-energy photon beams required relaxed DTA criteria for the brain IMRT QA, while low-energy photon beams showed better results even with tight DTA criteria.
- Is Part Of:
- Journal of radiotherapy in practice. Volume 18:Issue 2(2019)
- Journal:
- Journal of radiotherapy in practice
- Issue:
- Volume 18:Issue 2(2019)
- Issue Display:
- Volume 18, Issue 2 (2019)
- Year:
- 2019
- Volume:
- 18
- Issue:
- 2
- Issue Sort Value:
- 2019-0018-0002-0000
- Page Start:
- 127
- Page End:
- 131
- Publication Date:
- 2018-12-03
- Subjects:
- IMRT, -- Gafchromic EBT3 film, -- gamma analysis, -- quality assurance
Radiotherapy -- Periodicals
615.842005 - Journal URLs:
- http://journals.cambridge.org/action/displayJournal?jid=JRP ↗
- DOI:
- 10.1017/S1460396918000602 ↗
- Languages:
- English
- ISSNs:
- 1460-3969
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library HMNTS - ELD Digital store
- Ingest File:
- 13047.xml