Benchtop von Willebrand Factor Testing: Comparison of Commercially Available Ventricular Assist Devices and Evaluation of Variables for a Standardized Test Method. Issue 5 (July 2019)
- Record Type:
- Journal Article
- Title:
- Benchtop von Willebrand Factor Testing: Comparison of Commercially Available Ventricular Assist Devices and Evaluation of Variables for a Standardized Test Method. Issue 5 (July 2019)
- Main Title:
- Benchtop von Willebrand Factor Testing
- Authors:
- Coghill, Phillip A.
Kanchi, Suren
Azartash-Namin, Zheila J.
Long, James W.
Snyder, Trevor A. - Abstract:
- Abstract : Gastrointestinal bleeding occurs in 20–30% of patients receiving ventricular assist devices (VADs) due, in part, to acquired von Willebrand syndrome. We examined factors to optimize a benchtop method to quantify changes in von Willebrand Factor (VWF) multimer distribution and function in VADs, then applied them to evaluate commercially available devices. Human plasma was circulated through flow loops with VADs. Several experimental conditions were examined, including temperature, viscosity, and enzyme inhibition. Samples were analyzed for VWF collagen-binding activity (VWF:CB) and VWF antigen level. von Willebrand Factor multimer profiles were quantified using gel electrophoresis, near-infrared in-gel visualization, and densitometric analysis. The VWF:CB/antigen ratio in the HeartMate II, CentriMag, and HVAD exhibited average decreases of 46%, 44%, and 36% from baseline after 360 minutes of operation. High molecular weight (hVWF) multimer loss occurred within 30 minutes, although the Levacor and control loop profiles were unchanged. Varying temperature and viscosity altered hVWF degradation rate, but not the final results. Inhibition of a disintegrin and metalloprotease with thrombospondin type 1 repeats, member 13 (ADAMTS13) can potentially distinguish mechanoenzymatic cleavage of VWF from mechanical degradation. We developed a repeatable benchtop method to evaluate VWF compatibility of VADs similar to hemolysis testing that can be adopted for preclinical VADAbstract : Gastrointestinal bleeding occurs in 20–30% of patients receiving ventricular assist devices (VADs) due, in part, to acquired von Willebrand syndrome. We examined factors to optimize a benchtop method to quantify changes in von Willebrand Factor (VWF) multimer distribution and function in VADs, then applied them to evaluate commercially available devices. Human plasma was circulated through flow loops with VADs. Several experimental conditions were examined, including temperature, viscosity, and enzyme inhibition. Samples were analyzed for VWF collagen-binding activity (VWF:CB) and VWF antigen level. von Willebrand Factor multimer profiles were quantified using gel electrophoresis, near-infrared in-gel visualization, and densitometric analysis. The VWF:CB/antigen ratio in the HeartMate II, CentriMag, and HVAD exhibited average decreases of 46%, 44%, and 36% from baseline after 360 minutes of operation. High molecular weight (hVWF) multimer loss occurred within 30 minutes, although the Levacor and control loop profiles were unchanged. Varying temperature and viscosity altered hVWF degradation rate, but not the final results. Inhibition of a disintegrin and metalloprotease with thrombospondin type 1 repeats, member 13 (ADAMTS13) can potentially distinguish mechanoenzymatic cleavage of VWF from mechanical degradation. We developed a repeatable benchtop method to evaluate VWF compatibility of VADs similar to hemolysis testing that can be adopted for preclinical VAD evaluation. … (more)
- Is Part Of:
- ASAIO journal. Volume 65:Issue 5(2019)
- Journal:
- ASAIO journal
- Issue:
- Volume 65:Issue 5(2019)
- Issue Display:
- Volume 65, Issue 5 (2019)
- Year:
- 2019
- Volume:
- 65
- Issue:
- 5
- Issue Sort Value:
- 2019-0065-0005-0000
- Page Start:
- Page End:
- Publication Date:
- 2019-07
- Subjects:
- von Willebrand Factor -- hemocompatibility -- shear -- ventricular assist device -- blood pump
Artificial organs -- Periodicals
617 - Journal URLs:
- http://journals.lww.com/asaiojournal/pages/default.aspx ↗
http://journals.lww.com ↗ - DOI:
- 10.1097/MAT.0000000000000849 ↗
- Languages:
- English
- ISSNs:
- 1058-2916
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 1738.840500
British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 13042.xml