Feasibility of Attachable Ring Stimulator for Intraoperative Neuromonitoring during Thyroid Surgery. (30th January 2020)
- Record Type:
- Journal Article
- Title:
- Feasibility of Attachable Ring Stimulator for Intraoperative Neuromonitoring during Thyroid Surgery. (30th January 2020)
- Main Title:
- Feasibility of Attachable Ring Stimulator for Intraoperative Neuromonitoring during Thyroid Surgery
- Authors:
- Kim, Jongjin
Moon, Hyeon Jong
Chai, Young Jun
Lee, Jung-Man
Hwang, Ki-Tae
Wu, Che-Wei
Dionigi, Gianlorenzo
Kim, Hoon Yub
Park, Kyung Sik
Kim, Sang Wan
Yi, Ka Hee - Other Names:
- Horowitz Michael Academic Editor.
- Abstract:
- Abstract : Objective . Stimulator-attached dissecting instruments are useful for intraoperative nerve monitoring during thyroidectomy. The aim of this study was to evaluate the feasibility of an attachable ring stimulator (ARS) by comparing the electromyography (EMG) amplitudes evoked by an ARS and a conventional stimulator. Methods . Medical records of fourteen patients who underwent thyroidectomy using intraoperative neuromonitoring between June and August 2019 were retrospectively reviewed. The amplitudes of V1, R1, R2, and V2 signals were checked using both the ARS and a conventional stimulator, at the same point. Results . Both stimulators were tested on 20 recurrent laryngeal nerves (RLNs) and 20 vagus nerves (VNs). In all the nerves, the amplitudes of V1, R1, R2, and V2 were greater than 500 μ V. The mean amplitudes of V1, R1, R2, and V2 checked with the ARS were 1175, 1432, 1598, and 1279 μ V, respectively. The mean amplitudes of V1, R1, R2, and V2 checked with the conventional stimulator were 1140, 1425, 1557, and 1217 μ V, respectively. Difference between amplitudes evoked by the two stimulators for V1, R1, R2, and V2 was 77, 110, 102, and 99 μ V, respectively. There was no statistical difference in the amplitudes between the two groups for V1, R1, R2, and V2. Conclusion . The ARS transferred electric stimulation as effectively as the conventional stimulator. It is an effective tool for repeated stimulation and facilitates continuous feedback regarding theAbstract : Objective . Stimulator-attached dissecting instruments are useful for intraoperative nerve monitoring during thyroidectomy. The aim of this study was to evaluate the feasibility of an attachable ring stimulator (ARS) by comparing the electromyography (EMG) amplitudes evoked by an ARS and a conventional stimulator. Methods . Medical records of fourteen patients who underwent thyroidectomy using intraoperative neuromonitoring between June and August 2019 were retrospectively reviewed. The amplitudes of V1, R1, R2, and V2 signals were checked using both the ARS and a conventional stimulator, at the same point. Results . Both stimulators were tested on 20 recurrent laryngeal nerves (RLNs) and 20 vagus nerves (VNs). In all the nerves, the amplitudes of V1, R1, R2, and V2 were greater than 500 μ V. The mean amplitudes of V1, R1, R2, and V2 checked with the ARS were 1175, 1432, 1598, and 1279 μ V, respectively. The mean amplitudes of V1, R1, R2, and V2 checked with the conventional stimulator were 1140, 1425, 1557, and 1217 μ V, respectively. Difference between amplitudes evoked by the two stimulators for V1, R1, R2, and V2 was 77, 110, 102, and 99 μ V, respectively. There was no statistical difference in the amplitudes between the two groups for V1, R1, R2, and V2. Conclusion . The ARS transferred electric stimulation as effectively as the conventional stimulator. It is an effective tool for repeated stimulation and facilitates continuous feedback regarding the functional integrity of nerves during thyroid surgery. … (more)
- Is Part Of:
- International journal of endocrinology. Volume 2020(2020)
- Journal:
- International journal of endocrinology
- Issue:
- Volume 2020(2020)
- Issue Display:
- Volume 2020, Issue 2020 (2020)
- Year:
- 2020
- Volume:
- 2020
- Issue:
- 2020
- Issue Sort Value:
- 2020-2020-2020-0000
- Page Start:
- Page End:
- Publication Date:
- 2020-01-30
- Subjects:
- Endocrinology -- Periodicals
Endocrinology
Endocrinology -- Periodicals
Endocrine System Diseases -- Periodicals
Periodicals
616.4 - Journal URLs:
- https://www.hindawi.com/journals/ije/ ↗
http://bibpurl.oclc.org/web/41843 ↗
http://www.ncbi.nlm.nih.gov/pmc/journals/995/ ↗ - DOI:
- 10.1155/2020/5280939 ↗
- Languages:
- English
- ISSNs:
- 1687-8337
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library HMNTS - ELD Digital store
- Ingest File:
- 12775.xml