Optimization and characterization of a novel internally-cooled radiofrequency ablation system with optimized pulsing algorithm in an ex-vivo bovine liver. (1st October 2019)
- Record Type:
- Journal Article
- Title:
- Optimization and characterization of a novel internally-cooled radiofrequency ablation system with optimized pulsing algorithm in an ex-vivo bovine liver. (1st October 2019)
- Main Title:
- Optimization and characterization of a novel internally-cooled radiofrequency ablation system with optimized pulsing algorithm in an ex-vivo bovine liver
- Authors:
- Ben-David, Eliel
Nissenbaum, Isaac
Gurevich, Svetlana
Cosman, Eric R.
Goldberg, S. Nahum - Abstract:
- Abstract: Purpose: To prospectively characterize and optimize radiofrequency energy deposition to determine ideal parameters for achieving large ablation zones. Materials and methods: An internally-cooled RF system was used to perform 214 ablations in 72 ex-vivo bovine livers. Tip exposure (1–5 cm), electrode current (400–2500 mA), and application duration (3–15 min) were systematically varied. A pulsing algorithm optimized efficiency of RF deposition, including initial automatic ramping followed by adjustment in current, in response to changes in tissue impedance. Following the procedure ablation diameter and length were measured, sphericity calculated, and correlated with parameters of energy deposition and tissue temperatures. Results: Increasing electrode exposure from 1–5 cm produced linear increases in ablation diameter from 1.4 ± 0.1 to 5.3 ± 0.1 cm (y = 1.1x–0.5; R 2 = 0.93), and length (y = 1.18x + 0.34; R 2 = 0.92). A sphericity index >0.85 was noted at optimal energy setting for electrode exposures of 1–4 cm. Maximum temperatures post-ablation increased with active tip length from 68.5 ± 4.9 °C to 91.3 ± 1.5 °C in a logarithmic (y = 0.94ln(x)−2.75; R 2 = 0.90) or power relationship between temperature and the resultant ablation diameter (y = 0.27e 0.0295x ; R 2 = 0.76). A tight exponential relationship (y = 0.28x 0.38; R 2 = 0.97) was also observed between total energy deposition and ablation diameter. Finally, a multifactor relationship of the diameter ofAbstract: Purpose: To prospectively characterize and optimize radiofrequency energy deposition to determine ideal parameters for achieving large ablation zones. Materials and methods: An internally-cooled RF system was used to perform 214 ablations in 72 ex-vivo bovine livers. Tip exposure (1–5 cm), electrode current (400–2500 mA), and application duration (3–15 min) were systematically varied. A pulsing algorithm optimized efficiency of RF deposition, including initial automatic ramping followed by adjustment in current, in response to changes in tissue impedance. Following the procedure ablation diameter and length were measured, sphericity calculated, and correlated with parameters of energy deposition and tissue temperatures. Results: Increasing electrode exposure from 1–5 cm produced linear increases in ablation diameter from 1.4 ± 0.1 to 5.3 ± 0.1 cm (y = 1.1x–0.5; R 2 = 0.93), and length (y = 1.18x + 0.34; R 2 = 0.92). A sphericity index >0.85 was noted at optimal energy setting for electrode exposures of 1–4 cm. Maximum temperatures post-ablation increased with active tip length from 68.5 ± 4.9 °C to 91.3 ± 1.5 °C in a logarithmic (y = 0.94ln(x)−2.75; R 2 = 0.90) or power relationship between temperature and the resultant ablation diameter (y = 0.27e 0.0295x ; R 2 = 0.76). A tight exponential relationship (y = 0.28x 0.38; R 2 = 0.97) was also observed between total energy deposition and ablation diameter. Finally, a multifactor relationship of the diameter of ablation to electrode tip exposure and the time to first impedance rise was successfully modeled, with a root mean squared error of 1.9 mm and R 2 = 0.95. Conclusion: Large, reproducible, and spherical ablation areas can be achieved with the novel system described, with efficient delivery of RF energy deposited into tissue. These findings may have important clinical relevance in regards to the clinical utility of RF ablation compared to other competitive forms of thermal tumor ablation. … (more)
- Is Part Of:
- International journal of hyperthermia. Volume 36:Number 1(2020)
- Journal:
- International journal of hyperthermia
- Issue:
- Volume 36:Number 1(2020)
- Issue Display:
- Volume 36, Issue 1 (2020)
- Year:
- 2020
- Volume:
- 36
- Issue:
- 1
- Issue Sort Value:
- 2020-0036-0001-0000
- Page Start:
- 81
- Page End:
- 88
- Publication Date:
- 2019-10-01
- Subjects:
- Radiofrequency ablation -- internally-cooled -- optimization -- characterization -- modeling
Thermotherapy -- Periodicals
615.832 - Journal URLs:
- http://informahealthcare.com/loi/hth ↗
http://www.tandf.co.uk/journals/titles/02656736.asp ↗
http://informahealthcare.com ↗ - DOI:
- 10.1080/02656736.2019.1617901 ↗
- Languages:
- English
- ISSNs:
- 0265-6736
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 4542.297000
British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 12774.xml