Near-infrared spectroscopy-based methods for quantitative determination of active pharmaceutical ingredient in transdermal gel formulations. Issue 10 (26th November 2019)
- Record Type:
- Journal Article
- Title:
- Near-infrared spectroscopy-based methods for quantitative determination of active pharmaceutical ingredient in transdermal gel formulations. Issue 10 (26th November 2019)
- Main Title:
- Near-infrared spectroscopy-based methods for quantitative determination of active pharmaceutical ingredient in transdermal gel formulations
- Authors:
- Szabó, Éva
Gergely, Szilveszter
Spaits, Tamás
Simon, Tamás
Salgó, András - Abstract:
- Abstract: Semi-solid pharmaceutical forms, like topical and transdermal products, are widely used for medical purposes since they provide convenient and pain-free treatment for patients. In this study, two transdermal gel formulations with different active pharmaceutical ingredients were analyzed with near-infrared spectrophotometers with three distinct optical arrangements (dispersive, Fourier transform, and diode-array spectrophotometers) for the measurement of active pharmaceutical ingredient concentration. The reference concentrations were measured with high-performance liquid chromatography methods. The developing of calibration for reference concentrations based on near-infrared spectra was performed using partial least squares regression. The spectra from near-infrared spectrophotometer with diode-array optical arrangement presented the best root mean square error of cross-validation results with 0.045% and 0.144% for the two active pharmaceutical ingredients. The root mean square error of cross-validation using dispersive spectra were 0.125% and 0.193% and using Fourier transform spectra were 0.130% and 0.344% for the active pharmaceutical ingredients. The results of partial least squares regression were confirmed using simulated spectra based on the measured datasets. Near-infrared spectroscopy can be used successfully for the quantitative measurement of active pharmaceutical ingredient concentration in transdermal gel formulations. The type of optical arrangementsAbstract: Semi-solid pharmaceutical forms, like topical and transdermal products, are widely used for medical purposes since they provide convenient and pain-free treatment for patients. In this study, two transdermal gel formulations with different active pharmaceutical ingredients were analyzed with near-infrared spectrophotometers with three distinct optical arrangements (dispersive, Fourier transform, and diode-array spectrophotometers) for the measurement of active pharmaceutical ingredient concentration. The reference concentrations were measured with high-performance liquid chromatography methods. The developing of calibration for reference concentrations based on near-infrared spectra was performed using partial least squares regression. The spectra from near-infrared spectrophotometer with diode-array optical arrangement presented the best root mean square error of cross-validation results with 0.045% and 0.144% for the two active pharmaceutical ingredients. The root mean square error of cross-validation using dispersive spectra were 0.125% and 0.193% and using Fourier transform spectra were 0.130% and 0.344% for the active pharmaceutical ingredients. The results of partial least squares regression were confirmed using simulated spectra based on the measured datasets. Near-infrared spectroscopy can be used successfully for the quantitative measurement of active pharmaceutical ingredient concentration in transdermal gel formulations. The type of optical arrangements of spectrophotometers can influence the performance of calibrations and have to be optimized case by case. … (more)
- Is Part Of:
- Spectroscopy letters. Volume 52:Issue 10(2019)
- Journal:
- Spectroscopy letters
- Issue:
- Volume 52:Issue 10(2019)
- Issue Display:
- Volume 52, Issue 10 (2019)
- Year:
- 2019
- Volume:
- 52
- Issue:
- 10
- Issue Sort Value:
- 2019-0052-0010-0000
- Page Start:
- 599
- Page End:
- 611
- Publication Date:
- 2019-11-26
- Subjects:
- partial least squares regression -- transdermal gel formulations -- near-infrared spectroscopy -- simulated spectra
Spectrum analysis -- Periodicals
543.5 - Journal URLs:
- http://www.tandfonline.com/ ↗
- DOI:
- 10.1080/00387010.2019.1681459 ↗
- Languages:
- English
- ISSNs:
- 0038-7010
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 8411.120000
British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 12725.xml