Worldwide experience with a totally subcutaneous implantable defibrillator: early results from the EFFORTLESS S-ICD Registry. (26th March 2014)
- Record Type:
- Journal Article
- Title:
- Worldwide experience with a totally subcutaneous implantable defibrillator: early results from the EFFORTLESS S-ICD Registry. (26th March 2014)
- Main Title:
- Worldwide experience with a totally subcutaneous implantable defibrillator: early results from the EFFORTLESS S-ICD Registry
- Authors:
- Lambiase, Pier D.
Barr, Craig
Theuns, Dominic A.M.J.
Knops, Reinoud
Neuzil, Petr
Johansen, Jens Brock
Hood, Margaret
Pedersen, Susanne
Kääb, Stefan
Murgatroyd, Francis
Reeve, Helen L.
Carter, Nathan
Boersma, Lucas - Other Names:
- Adragão P. author non-byline.
Agarwal S. author non-byline.
Barr C. author non-byline.
Boersma L. author non-byline.
Brock-Johanssen J. author non-byline.
Butter C. author non-byline.
Calò L. author non-byline.
Eckhardt L. author non-byline.
Gulizia M. author non-byline.
Scholten M. author non-byline.
Dekker L. author non-byline.
Khiani R. author non-byline.
Hjortshot S. author non-byline.
Høgh Petersen H. author non-byline.
Hood M. author non-byline.
Kääb S. author non-byline.
Knops R. author non-byline.
Kuschyk J. author non-byline.
Lambiase P. author non-byline.
Maass K. A. author non-byline.
McLeod K. author non-byline.
Molon G. author non-byline.
Morgan J. author non-byline.
Mortensen P. author non-byline.
Murgatroyd F. author non-byline.
Neuzil P. author non-byline.
Pepper C. author non-byline.
Sheridan P. author non-byline.
Stellbrink C. author non-byline.
Stuart G. author non-byline.
Theuns D. author non-byline.
Vernooy K. author non-byline.
Veltmann C. author non-byline.
Wende C. author non-byline.
… (more) - Abstract:
- Abstract : Aims: The totally subcutaneous implantable-defibrillator (S-ICD) is a new alternative to the conventional transvenous ICD system to minimize intravascular lead complications. There are limited data describing the long-term performance of the S-ICD. This paper presents the first large international patient population collected as part of the EFFORTLESS S-ICD Registry. Methods and results: The EFFORTLESS S-ICD Registry is a non-randomized, standard of care, multicentre Registry designed to collect long-term, system-related, clinical, and patient reported outcome data from S-ICD implanted patients since June 2009. Follow-up data are systematically collected over 60-month post-implant including Quality of Life. The study population of 472 patients of which 241 (51%) were enrolled prospectively has a mean follow-up duration of 558 days (range 13–1342 days, median 498 days), 72% male, mean age of 49 ± 18 years (range 9–88 years), 42% mean left ventricular ejection fraction. Complication-free rates were 97 and 94%, at 30 and 360 days, respectively. Three hundred and seventeen spontaneous episodes were recorded in 85 patients during the follow-up period. Of these episodes, 169 (53%) received therapy, 93 being for Ventricular Tachycardia/Fibrillation (VT/VF). One patient died of recurrent VF and severe bradycardia. Regarding discrete VT/VF episodes, first shock conversion efficacy was 88% with 100% overall successful clinical conversion after a maximum of five shocks. TheAbstract : Aims: The totally subcutaneous implantable-defibrillator (S-ICD) is a new alternative to the conventional transvenous ICD system to minimize intravascular lead complications. There are limited data describing the long-term performance of the S-ICD. This paper presents the first large international patient population collected as part of the EFFORTLESS S-ICD Registry. Methods and results: The EFFORTLESS S-ICD Registry is a non-randomized, standard of care, multicentre Registry designed to collect long-term, system-related, clinical, and patient reported outcome data from S-ICD implanted patients since June 2009. Follow-up data are systematically collected over 60-month post-implant including Quality of Life. The study population of 472 patients of which 241 (51%) were enrolled prospectively has a mean follow-up duration of 558 days (range 13–1342 days, median 498 days), 72% male, mean age of 49 ± 18 years (range 9–88 years), 42% mean left ventricular ejection fraction. Complication-free rates were 97 and 94%, at 30 and 360 days, respectively. Three hundred and seventeen spontaneous episodes were recorded in 85 patients during the follow-up period. Of these episodes, 169 (53%) received therapy, 93 being for Ventricular Tachycardia/Fibrillation (VT/VF). One patient died of recurrent VF and severe bradycardia. Regarding discrete VT/VF episodes, first shock conversion efficacy was 88% with 100% overall successful clinical conversion after a maximum of five shocks. The 360-day inappropriate shock rate was 7% with the vast majority occurring for oversensing (62/73 episodes), primarily of cardiac signals (94% of oversensed episodes). Conclusion: The first large cohort of real-world data from an International patient S-ICD population demonstrates appropriate system performance with clinical event rates and inappropriate shock rates comparable with those reported for conventional ICDs. Clinical trial registration URL: http://www.clinicaltrials.gov . Unique identifier NCT01085435. … (more)
- Is Part Of:
- European heart journal. Volume 35:Number 25(2014:Jul.)
- Journal:
- European heart journal
- Issue:
- Volume 35:Number 25(2014:Jul.)
- Issue Display:
- Volume 35, Issue 25 (2014)
- Year:
- 2014
- Volume:
- 35
- Issue:
- 25
- Issue Sort Value:
- 2014-0035-0025-0000
- Page Start:
- 1657
- Page End:
- 1665
- Publication Date:
- 2014-03-26
- Subjects:
- Subcutaneous ICD -- Ventricular arrhythmias -- Cardiac arrest -- Primary prevention -- Secondary prevention
Cardiology -- Periodicals
Heart -- Diseases -- Periodicals
616.12005 - Journal URLs:
- http://eurheartj.oxfordjournals.org/ ↗
http://ukcatalogue.oup.com/ ↗ - DOI:
- 10.1093/eurheartj/ehu112 ↗
- Languages:
- English
- ISSNs:
- 0195-668X
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 3829.717500
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 12693.xml