Quantitation of estradiol by competitive light‐initiated chemiluminescent assay using estriol as competitive antigen. Issue 1 (24th August 2019)
- Record Type:
- Journal Article
- Title:
- Quantitation of estradiol by competitive light‐initiated chemiluminescent assay using estriol as competitive antigen. Issue 1 (24th August 2019)
- Main Title:
- Quantitation of estradiol by competitive light‐initiated chemiluminescent assay using estriol as competitive antigen
- Authors:
- Li, Jiuzhi
Li, Liuxu
Bian, Ying
Yu, Yang
Qiang, Zhonghua
Zhang, Yuexiang
Li, Huiqiang - Abstract:
- Abstract: Background: Light‐initiated chemiluminescent assays (LICA) are homogeneous assays that are sensitive, specific, and free of separation and washing steps and have high throughput and high precision. Methods: In this research, we developed a competitive method by LICA to achieve accurate quantification of estradiol (E2) in human serum. E2 competed with estriol (E3) for binding to anti‐human E2 antibodies. E3 was linked to biotin via bovine serum albumin as a linker. As this assay used competition between the labeled tracer and the analyte, an increase in E2 concentration will cause a signal decrease. Results: The expected detection range of E2 was 20‐5000 pg/mL. The analytical and functional sensitivities were 7.16 and 13.7 pg/mL, respectively. The intra‐ and inter‐assay coefficients of variation were both below 15%, and the recovery rate ranged from 97.5% to 106.8%. The interference rates ranged from −3.6% to 5.4% and met detection requirements for E2 in hyperbilirubinemia, hemolysis, and lipemia in clinical samples. In addition, the cross‐reactivity rates between E2 and structural analogs and some reproductive hormones varied from 1.9% to 10.6% which showed that LICA is highly specific for E2. Moreover, our results showed high accordance with the IMMULITE 2000 ( y = 0.6695 x + 47.92, r 2 = .843) and VIDAS systems ( y = 1.099 x − 821.5, r 2 = .9392). Conclusion: Our data show that the LICA, which is easy to automate, is a promising technique for quantificationAbstract: Background: Light‐initiated chemiluminescent assays (LICA) are homogeneous assays that are sensitive, specific, and free of separation and washing steps and have high throughput and high precision. Methods: In this research, we developed a competitive method by LICA to achieve accurate quantification of estradiol (E2) in human serum. E2 competed with estriol (E3) for binding to anti‐human E2 antibodies. E3 was linked to biotin via bovine serum albumin as a linker. As this assay used competition between the labeled tracer and the analyte, an increase in E2 concentration will cause a signal decrease. Results: The expected detection range of E2 was 20‐5000 pg/mL. The analytical and functional sensitivities were 7.16 and 13.7 pg/mL, respectively. The intra‐ and inter‐assay coefficients of variation were both below 15%, and the recovery rate ranged from 97.5% to 106.8%. The interference rates ranged from −3.6% to 5.4% and met detection requirements for E2 in hyperbilirubinemia, hemolysis, and lipemia in clinical samples. In addition, the cross‐reactivity rates between E2 and structural analogs and some reproductive hormones varied from 1.9% to 10.6% which showed that LICA is highly specific for E2. Moreover, our results showed high accordance with the IMMULITE 2000 ( y = 0.6695 x + 47.92, r 2 = .843) and VIDAS systems ( y = 1.099 x − 821.5, r 2 = .9392). Conclusion: Our data show that the LICA, which is easy to automate, is a promising technique for quantification of E2 in human serum and could be used for clinical detection. … (more)
- Is Part Of:
- Journal of clinical laboratory analysis. Volume 34:Issue 1(2020)
- Journal:
- Journal of clinical laboratory analysis
- Issue:
- Volume 34:Issue 1(2020)
- Issue Display:
- Volume 34, Issue 1 (2020)
- Year:
- 2020
- Volume:
- 34
- Issue:
- 1
- Issue Sort Value:
- 2020-0034-0001-0000
- Page Start:
- n/a
- Page End:
- n/a
- Publication Date:
- 2019-08-24
- Subjects:
- equilibrium competitive assay -- estradiol -- estriol -- light‐initiated chemiluminescent assay -- quantitative analysis
Diagnosis, Laboratory -- Periodicals
Medical laboratory technology -- Periodicals
616 - Journal URLs:
- http://onlinelibrary.wiley.com/ ↗
- DOI:
- 10.1002/jcla.23014 ↗
- Languages:
- English
- ISSNs:
- 0887-8013
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 4958.520000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 12625.xml