Biomechanical Evaluation of Proximal Hamstring Repair: All-Suture Anchor Versus Titanium Suture Anchor. Issue 1 (30th December 2019)
- Record Type:
- Journal Article
- Title:
- Biomechanical Evaluation of Proximal Hamstring Repair: All-Suture Anchor Versus Titanium Suture Anchor. Issue 1 (30th December 2019)
- Main Title:
- Biomechanical Evaluation of Proximal Hamstring Repair: All-Suture Anchor Versus Titanium Suture Anchor
- Authors:
- Otto, Alexander
DiCosmo, Alyssa M.
Baldino, Joshua B.
Mehl, Julian
Obopilwe, Elifho
Cote, Mark P.
Imhoff, Andreas B.
Beitzel, Knut
Mazzocca, Augustus D.
Coyner, Katherine - Abstract:
- Background: Proximal hamstring avulsions are severe tendon injuries and are commonly sports-related. Open and endoscopic techniques as well as different anchor configurations have already been described for proximal hamstring repair. Novel all-suture anchors have been developed to provide decreased bone loss during placement and reduced occupied bone volume when compared with titanium suture anchors. Hypothesis: Complete proximal hamstring avulsions repaired with all-suture anchors will demonstrate equal load to failure and comparable displacement under cyclic loading when compared with titanium suture anchors. Study Design: Controlled laboratory study. Methods: Complete proximal hamstring avulsions were created in 18 paired cadaveric specimens (mean ± SD age, 63.0 ± 10.4 years). Either all-suture anchors or titanium suture anchors were used for repair. Cyclic loading from 10 to 125 N at 1 Hz was performed for 1500 cycles with a material testing machine. Displacement was assessed along anterior and posterior aspects of the tendon repair with optical tracking. Specimens were loaded to failure at a rate of 120 mm/min. Displacement, load to failure, and repair construct stiffness were compared between matched pairs with the Wilcoxon signed-rank test. Correlations were determined by Spearman rho analysis. Results: The all-suture anchors showed significantly higher load-to-failure values when compared with the titanium anchor repairs (799.64 ± 257.1 vs 573.27 ± 89.9 N; P = .008).Background: Proximal hamstring avulsions are severe tendon injuries and are commonly sports-related. Open and endoscopic techniques as well as different anchor configurations have already been described for proximal hamstring repair. Novel all-suture anchors have been developed to provide decreased bone loss during placement and reduced occupied bone volume when compared with titanium suture anchors. Hypothesis: Complete proximal hamstring avulsions repaired with all-suture anchors will demonstrate equal load to failure and comparable displacement under cyclic loading when compared with titanium suture anchors. Study Design: Controlled laboratory study. Methods: Complete proximal hamstring avulsions were created in 18 paired cadaveric specimens (mean ± SD age, 63.0 ± 10.4 years). Either all-suture anchors or titanium suture anchors were used for repair. Cyclic loading from 10 to 125 N at 1 Hz was performed for 1500 cycles with a material testing machine. Displacement was assessed along anterior and posterior aspects of the tendon repair with optical tracking. Specimens were loaded to failure at a rate of 120 mm/min. Displacement, load to failure, and repair construct stiffness were compared between matched pairs with the Wilcoxon signed-rank test. Correlations were determined by Spearman rho analysis. Results: The all-suture anchors showed significantly higher load-to-failure values when compared with the titanium anchor repairs (799.64 ± 257.1 vs 573.27 ± 89.9 N; P = .008). There was no significant difference in displacement between all-suture anchors and titanium suture anchors at the anterior aspect (6.60 ± 2.2 vs 5.49 ± 1.1 mm; P = .26) or posterior aspect (5.87 ± 2.08 vs 5.23 ± 1.37 mm; P = .678) of the repaired hamstring tendons. Conclusion: All-suture anchors demonstrated similar displacement and superior load to failure when compared with titanium suture anchors. Clinical Relevance: The results of this study suggest that all-suture anchors are an equivalent alternative to titanium suture anchors for proximal hamstring avulsion repair. … (more)
- Is Part Of:
- Orthopaedic journal of sports medicine. Volume 8:Issue 1(2020)
- Journal:
- Orthopaedic journal of sports medicine
- Issue:
- Volume 8:Issue 1(2020)
- Issue Display:
- Volume 8, Issue 1 (2020)
- Year:
- 2020
- Volume:
- 8
- Issue:
- 1
- Issue Sort Value:
- 2020-0008-0001-0000
- Page Start:
- Page End:
- Publication Date:
- 2019-12-30
- Subjects:
- proximal hamstring avulsion -- hip -- all-suture anchor -- titanium suture anchor -- biomechanics -- proximal hamstring repair
Sports medicine -- Periodicals
Orthopedics -- Periodicals
Arthroscopy -- Periodicals
Arthroplasty -- Periodicals
Knee -- Surgery -- Periodicals
616.7 - Journal URLs:
- http://www.sagepublications.com/ ↗
- DOI:
- 10.1177/2325967119892925 ↗
- Languages:
- English
- ISSNs:
- 2325-9671
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 12601.xml