An unusual cause of apparent rebreathing associated with anaesthesia delivered by a coaxial circle breathing system. Issue 2 (6th December 2016)
- Record Type:
- Journal Article
- Title:
- An unusual cause of apparent rebreathing associated with anaesthesia delivered by a coaxial circle breathing system. Issue 2 (6th December 2016)
- Main Title:
- An unusual cause of apparent rebreathing associated with anaesthesia delivered by a coaxial circle breathing system
- Authors:
- Goosen, Leonard
Hall, Alexander J
Elsey, Chris
Uncles, David R - Abstract:
- Summary: This case report describes an unusual case of equipment failure resulting in the appearance of a rebreathing artefact on an associated capnography waveform trace that we believe has not been previously reported. The rebreathing artefact occurred with a single‐use coaxial circle breathing system which included an integral gas sampling line within the expiratory limb. Close examination of the system revealed a defect in the sampling line resulting in gas sampling predominantly from the expiratory limb of the circle system. This defect was not detectable by either the anaesthetic machine automated self‐testing cycle or visual inspection of the breathing system.
- Is Part Of:
- Anaesthesia cases. Volume 4:Issue 2(2016)
- Journal:
- Anaesthesia cases
- Issue:
- Volume 4:Issue 2(2016)
- Issue Display:
- Volume 4, Issue 2 (2016)
- Year:
- 2016
- Volume:
- 4
- Issue:
- 2
- Issue Sort Value:
- 2016-0004-0002-0000
- Page Start:
- 132
- Page End:
- 135
- Publication Date:
- 2016-12-06
- Subjects:
- 617.9605
- Journal URLs:
- http://www.anaesthesiacases.org/ ↗
http://onlinelibrary.wiley.com/ ↗ - DOI:
- 10.21466/ac.AUCOARA.2016 ↗
- Languages:
- English
- ISSNs:
- 2396-8397
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 12598.xml