Beam angle optimization using angular dependency of range variation assessed via water equivalent path length (WEPL) calculation for head and neck proton therapy. (January 2020)
- Record Type:
- Journal Article
- Title:
- Beam angle optimization using angular dependency of range variation assessed via water equivalent path length (WEPL) calculation for head and neck proton therapy. (January 2020)
- Main Title:
- Beam angle optimization using angular dependency of range variation assessed via water equivalent path length (WEPL) calculation for head and neck proton therapy
- Authors:
- Kim, Jihun
Park, Yang-Kyun
Sharp, Gregory
Busse, Paul
Winey, Brian - Abstract:
- Highlights: For posterior oblique angles, proton range variation is most sensitive to anatomic change in patients and patient setup error. Angular dependency of proton range variation was observed in 11 head and neck patients. Dosimetric impact of proton range variation may be reduced by selecting an optimal beam configuration. Calculating WEPL on scatter-corrected CBCT is a viable option to evaluate proton range variation. Abstract: Purpose: To investigate angular sensitivity of proton range variation due to anatomic change in patients and patient setup error via water equivalent path length (WEPL) calculations. Methods: Proton range was estimated by calculating WEPL to the distal edge of target volume using planning CT (pCT) and weekly scatter-corrected cone-beam CT (CBCT) images of 11 head and neck patients. Range variation was estimated as the difference between the distal WEPLs calculated on pCT and scatter-corrected CBCT (cCBCT). This WEPL analysis was performed every five degrees ipsilaterally to the target. Statistics of the distal WEPL difference were calculated over the distal area to compare between different beam angles. Physician-defined contours were used for the WEPL calculation on both pCT and cCBCT, not considering local deformation of target volume. It was also tested if a couch kick (10°) can mitigate the range variation due to anatomic change and patient setup error. Results: For most of the patients considered, median, 75% quantile, and 95% quantile ofHighlights: For posterior oblique angles, proton range variation is most sensitive to anatomic change in patients and patient setup error. Angular dependency of proton range variation was observed in 11 head and neck patients. Dosimetric impact of proton range variation may be reduced by selecting an optimal beam configuration. Calculating WEPL on scatter-corrected CBCT is a viable option to evaluate proton range variation. Abstract: Purpose: To investigate angular sensitivity of proton range variation due to anatomic change in patients and patient setup error via water equivalent path length (WEPL) calculations. Methods: Proton range was estimated by calculating WEPL to the distal edge of target volume using planning CT (pCT) and weekly scatter-corrected cone-beam CT (CBCT) images of 11 head and neck patients. Range variation was estimated as the difference between the distal WEPLs calculated on pCT and scatter-corrected CBCT (cCBCT). This WEPL analysis was performed every five degrees ipsilaterally to the target. Statistics of the distal WEPL difference were calculated over the distal area to compare between different beam angles. Physician-defined contours were used for the WEPL calculation on both pCT and cCBCT, not considering local deformation of target volume. It was also tested if a couch kick (10°) can mitigate the range variation due to anatomic change and patient setup error. Results: For most of the patients considered, median, 75% quantile, and 95% quantile of the distal WEPL difference were largest for posterior oblique angles, indicating a higher chance of overdosing normal tissues at distal edge with these angles. Using a couch kick resulted in decrease in the WEPL difference for some posterior oblique angles. Conclusions: It was demonstrated that the WEPL change has angular dependency for the cohort of head and neck cancer patients. Selecting beam configuration robust to anatomic change in patient and patient setup error may improve the treatment outcome of head and neck proton therapy. … (more)
- Is Part Of:
- Physica medica. Volume 69(2020)
- Journal:
- Physica medica
- Issue:
- Volume 69(2020)
- Issue Display:
- Volume 69, Issue 2020 (2020)
- Year:
- 2020
- Volume:
- 69
- Issue:
- 2020
- Issue Sort Value:
- 2020-0069-2020-0000
- Page Start:
- 19
- Page End:
- 27
- Publication Date:
- 2020-01
- Subjects:
- Water equivalent path length -- Angular dependency -- Proton therapy -- Anatomic change -- Patient setup error -- Cone-beam CT -- Scatter correction
Medical physics -- Periodicals
Biophysics -- Periodicals
Biophysics -- Periodicals
Imagerie médicale -- Périodiques
Radiothérapie -- Périodiques
Rayons X -- Sécurité -- Mesures -- Périodiques
Physique -- Périodiques
Médecine -- Périodiques
610.153 - Journal URLs:
- http://www.sciencedirect.com/science/journal/11201797 ↗
http://www.clinicalkey.com/dura/browse/journalIssue/11201797 ↗
http://www.clinicalkey.com.au/dura/browse/journalIssue/11201797 ↗
http://www.elsevier.com/journals ↗
http://www.physicamedica.com ↗ - DOI:
- 10.1016/j.ejmp.2019.11.021 ↗
- Languages:
- English
- ISSNs:
- 1120-1797
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 6475.070000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 12565.xml