Performance of the Linox implantable cardioverter defibrillator leads: A single‐center experience. Issue 12 (30th October 2019)
- Record Type:
- Journal Article
- Title:
- Performance of the Linox implantable cardioverter defibrillator leads: A single‐center experience. Issue 12 (30th October 2019)
- Main Title:
- Performance of the Linox implantable cardioverter defibrillator leads: A single‐center experience
- Authors:
- Marai, Ibrahim
Milman, Anat
Nof, Eyal
Gurevitz, Osnat
Barlev, David
Lipchenca, Igor
Bachar, Sharona
Glikson, Michael
Beinart, Roy - Abstract:
- Abstract: Background: An early failure of Biotronik Linox implantable cardioverter defibrillator (ICD) leads has been reported from several centers. Aim: To compare the performance of Linox ICD leads with different other ICD leads as a report of the Sheba Medical Center experience. Methods: All patients who had implantation of Linox ICD leads between 2007 and 2016 were included in this study. ICD lead failure was defined as low‐ or high‐voltage impedance; failure to capture, sense, or defibrillate; or the presence of nonphysiological signals not due to external interference. The survival probability of Linox leads was determined and compared to Medtronic Sprint Quattro ICD leads. Results: A total of 340 patients (age 64.4 ± 1.8 years) were included in this analysis. They were followed up to 105 months (mean 45.7 ± 7, median 44 (Interquartile range (IQR) 26‐63) months). Twelve patients (3.5%) met the criteria for lead failure within 61.2 ± 22.9 months (median 66.5 [IQR 48–85 months]) post implantation. Noise with inappropriate ventricular arrhythmias detection, with or without therapy, was seen in 10 patients (83%). High pacing thresholds and high impedances were detected in two patients (17%). The survival probability of Linox leads at 60 months (97.3%) was similar to the survival probability of Sprint Quattro leads (98.2%) ( P = .58). Nevertheless, the survival probability at 105 months was much lower (81% vs 97%, Linox ICD lead and Sprint Quattro lead, respectively, PAbstract: Background: An early failure of Biotronik Linox implantable cardioverter defibrillator (ICD) leads has been reported from several centers. Aim: To compare the performance of Linox ICD leads with different other ICD leads as a report of the Sheba Medical Center experience. Methods: All patients who had implantation of Linox ICD leads between 2007 and 2016 were included in this study. ICD lead failure was defined as low‐ or high‐voltage impedance; failure to capture, sense, or defibrillate; or the presence of nonphysiological signals not due to external interference. The survival probability of Linox leads was determined and compared to Medtronic Sprint Quattro ICD leads. Results: A total of 340 patients (age 64.4 ± 1.8 years) were included in this analysis. They were followed up to 105 months (mean 45.7 ± 7, median 44 (Interquartile range (IQR) 26‐63) months). Twelve patients (3.5%) met the criteria for lead failure within 61.2 ± 22.9 months (median 66.5 [IQR 48–85 months]) post implantation. Noise with inappropriate ventricular arrhythmias detection, with or without therapy, was seen in 10 patients (83%). High pacing thresholds and high impedances were detected in two patients (17%). The survival probability of Linox leads at 60 months (97.3%) was similar to the survival probability of Sprint Quattro leads (98.2%) ( P = .58). Nevertheless, the survival probability at 105 months was much lower (81% vs 97%, Linox ICD lead and Sprint Quattro lead, respectively, P = .0039). Conclusion: Linox ICD leads have higher late failure rates compared to Sprint Quattro leads. These findings need to be confirmed in larger scale studies. … (more)
- Is Part Of:
- Pacing and clinical electrophysiology. Volume 42:Issue 12(2019)
- Journal:
- Pacing and clinical electrophysiology
- Issue:
- Volume 42:Issue 12(2019)
- Issue Display:
- Volume 42, Issue 12 (2019)
- Year:
- 2019
- Volume:
- 42
- Issue:
- 12
- Issue Sort Value:
- 2019-0042-0012-0000
- Page Start:
- 1524
- Page End:
- 1528
- Publication Date:
- 2019-10-30
- Subjects:
- lead failure -- Linox -- Sprint Quattro
Cardiac pacing -- Periodicals
Electrophysiology -- Periodicals
617.4120645 - Journal URLs:
- http://onlinelibrary.wiley.com/journal/10.1111/(ISSN)1540-8159 ↗
http://www.blackwell-synergy.com/rd.asp?goto=journal&code=pace ↗
http://www.futuraco.com/journalsf.htm ↗
http://onlinelibrary.wiley.com/ ↗
http://firstsearch.oclc.org ↗
http://firstsearch.oclc.org/journal=0147-8389;screen=info;ECOIP ↗ - DOI:
- 10.1111/pace.13816 ↗
- Languages:
- English
- ISSNs:
- 0147-8389
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 6328.210000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 12477.xml