Technical Note: MR‐visualization of interventional devices using transient field alterations and balanced steady‐state free precession imaging. Issue 11 (20th October 2015)
- Record Type:
- Journal Article
- Title:
- Technical Note: MR‐visualization of interventional devices using transient field alterations and balanced steady‐state free precession imaging. Issue 11 (20th October 2015)
- Main Title:
- Technical Note: MR‐visualization of interventional devices using transient field alterations and balanced steady‐state free precession imaging
- Authors:
- Eibofner, Frank
Martirosian, Petros
Würslin, Christian
Graf, Hansjörg
Syha, Roland
Clasen, Stephan - Abstract:
- Abstract : Purpose: In interventional magnetic resonance imaging, instruments can be equipped with conducting wires for visualization by current application. The potential of sequence triggered application of transient direct currents in balanced steady‐state free precession (bSSFP) imaging is demonstrated. Methods: A conductor and a modified catheter were examined in water phantoms and in an ex vivo porcine liver. The current was switched by a trigger pulse in the bSSFP sequence in an interval between radiofrequency pulse and signal acquisition. Magnitude and phase images were recorded. Regions with transient field alterations were evaluated by a postprocessing algorithm. A phase mask was computed and overlaid with the magnitude image. Results: Transient field alterations caused continuous phase shifts, which were separated by the postprocessing algorithm from phase jumps due to persistent field alterations. The overlaid images revealed the position of the conductor. The modified catheter generated visible phase offset in all orientations toward the static magnetic field and could be unambiguously localized in the ex vivo porcine liver. Conclusions: The application of a sequence triggered, direct current in combination with phase imaging allows conspicuous localization of interventional devices with a bSSFP sequence.
- Is Part Of:
- Medical physics. Volume 42:Issue 11(2015)
- Journal:
- Medical physics
- Issue:
- Volume 42:Issue 11(2015)
- Issue Display:
- Volume 42, Issue 11 (2015)
- Year:
- 2015
- Volume:
- 42
- Issue:
- 11
- Issue Sort Value:
- 2015-0042-0011-0000
- Page Start:
- 6558
- Page End:
- 6563
- Publication Date:
- 2015-10-20
- Subjects:
- biomedical MRI -- liver -- phantoms
Clinical applications
Involving electronic [emr] or nuclear [nmr] magnetic resonance, e.g. magnetic resonance imaging -- Biological material, e.g. blood, urine; Haemocytometers
MR -- interventional MRI -- image artifacts -- bSSFP -- phase imaging
Direct current power transmission -- Liver -- Static magnetic fields -- Electric currents -- Magnetic resonance imaging -- Medical image artifacts -- Titanium -- Bioelectrochemistry
Medical physics -- Periodicals
Medical physics
Geneeskunde
Natuurkunde
Toepassingen
Biophysics
Periodicals
Periodicals
Electronic journals
610.153 - Journal URLs:
- http://scitation.aip.org/content/aapm/journal/medphys ↗
https://aapm.onlinelibrary.wiley.com/journal/24734209 ↗
http://www.aip.org/ ↗ - DOI:
- 10.1118/1.4932629 ↗
- Languages:
- English
- ISSNs:
- 0094-2405
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 5531.130000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 12410.xml