Technical Note: Motion‐perturbation method applied to dosimetry of dynamic MLC target tracking—A proof‐of‐concept. Issue 11 (7th October 2015)
- Record Type:
- Journal Article
- Title:
- Technical Note: Motion‐perturbation method applied to dosimetry of dynamic MLC target tracking—A proof‐of‐concept. Issue 11 (7th October 2015)
- Main Title:
- Technical Note: Motion‐perturbation method applied to dosimetry of dynamic MLC target tracking—A proof‐of‐concept
- Authors:
- Feygelman, Vladimir
Tonner, Brian
Stambaugh, Cassandra
Hunt, Dylan
Zhang, Geoffrey
Moros, Eduardo
Nelms, Benjamin E. - Abstract:
- Abstract : Purpose: Previous studies show that dose to a moving target can be estimated using 4D measurement‐guided dose reconstruction based on a process called virtual motion simulation, or VMS. A potential extension of VMS is to estimate dose during dynamic multileaf collimator (MLC)‐tracking treatments. The authors introduce a modified VMS method and quantify its performance as proof‐of‐concept for tracking applications. Methods: Direct measurements with a moving biplanar diode array were used to verify accuracy of the VMS dose estimates. A tracking environment for variably sized circular MLC apertures was simulated by sending preprogrammed control points to the MLC while simultaneously moving the accelerator treatment table. Sensitivity of the method to simulated tracking latency (0–700 ms) was also studied. Potential applicability of VMS to fast changing beam apertures was evaluated by modeling, based on the demonstrated dependence of the cumulative dose on the temporal dose gradient. Results: When physical and virtual latencies were matched, the agreement rates (2% global/2 mm gamma) between the VMS and the biplanar dosimeter were above 96%. When compared to their own reference dose (0 induced latency), the agreement rates for VMS and biplanar array track closely up to 200 ms of induced latency with 10% low‐dose cutoff threshold and 300 ms with 50% cutoff. Time‐resolved measurements suggest that even in the modulated beams, the error in the cumulative dose introducedAbstract : Purpose: Previous studies show that dose to a moving target can be estimated using 4D measurement‐guided dose reconstruction based on a process called virtual motion simulation, or VMS. A potential extension of VMS is to estimate dose during dynamic multileaf collimator (MLC)‐tracking treatments. The authors introduce a modified VMS method and quantify its performance as proof‐of‐concept for tracking applications. Methods: Direct measurements with a moving biplanar diode array were used to verify accuracy of the VMS dose estimates. A tracking environment for variably sized circular MLC apertures was simulated by sending preprogrammed control points to the MLC while simultaneously moving the accelerator treatment table. Sensitivity of the method to simulated tracking latency (0–700 ms) was also studied. Potential applicability of VMS to fast changing beam apertures was evaluated by modeling, based on the demonstrated dependence of the cumulative dose on the temporal dose gradient. Results: When physical and virtual latencies were matched, the agreement rates (2% global/2 mm gamma) between the VMS and the biplanar dosimeter were above 96%. When compared to their own reference dose (0 induced latency), the agreement rates for VMS and biplanar array track closely up to 200 ms of induced latency with 10% low‐dose cutoff threshold and 300 ms with 50% cutoff. Time‐resolved measurements suggest that even in the modulated beams, the error in the cumulative dose introduced by the 200 ms VMS time resolution is not likely to exceed 0.5%. Conclusions: Based on current results and prior benchmarks of VMS accuracy, the authors postulate that this approach should be applicable to any MLC‐tracking treatments where leaf speeds do not exceed those of the current Varian accelerators. … (more)
- Is Part Of:
- Medical physics. Volume 42:Issue 11(2015)
- Journal:
- Medical physics
- Issue:
- Volume 42:Issue 11(2015)
- Issue Display:
- Volume 42, Issue 11 (2015)
- Year:
- 2015
- Volume:
- 42
- Issue:
- 11
- Issue Sort Value:
- 2015-0042-0011-0000
- Page Start:
- 6147
- Page End:
- 6151
- Publication Date:
- 2015-10-07
- Subjects:
- cancer -- collimators -- dosimeters -- dosimetry -- radiation therapy -- tumours
Dose‐volume analysis -- Collimators -- Cancer
Radiation therapy -- Dosimeters -- Using diaphragms, collimators -- Scintigraphy
dynamic target tracking -- tumor motion -- radiation dosimetry -- MLC -- dose reconstruction
Multileaf collimators -- Dosimetry -- Kinematics -- Computer software -- Real time information delivery -- Cancer -- Three dimensional image processing -- Tracking devices -- Time measurement -- Electric measurements
Medical physics -- Periodicals
Medical physics
Geneeskunde
Natuurkunde
Toepassingen
Biophysics
Periodicals
Periodicals
Electronic journals
610.153 - Journal URLs:
- http://scitation.aip.org/content/aapm/journal/medphys ↗
https://aapm.onlinelibrary.wiley.com/journal/24734209 ↗
http://www.aip.org/ ↗ - DOI:
- 10.1118/1.4931605 ↗
- Languages:
- English
- ISSNs:
- 0094-2405
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 5531.130000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 12410.xml