Evaluation of Image Quality Improvements When Adding Patient Outline Constraints into a Generalized Scatter PET Reconstruction Algorithm. (2nd May 2013)
- Record Type:
- Journal Article
- Title:
- Evaluation of Image Quality Improvements When Adding Patient Outline Constraints into a Generalized Scatter PET Reconstruction Algorithm. (2nd May 2013)
- Main Title:
- Evaluation of Image Quality Improvements When Adding Patient Outline Constraints into a Generalized Scatter PET Reconstruction Algorithm
- Authors:
- Sun, Hongyan
Pistorius, Stephen - Other Names:
- Anashkin E. Academic Editor.
Han W. Academic Editor. - Abstract:
- Abstract : Scattered coincidences degrade image contrast and compromise quantitative accuracy in positron emission tomography (PET). A number of approaches to estimating and correcting scattered coincidences have been proposed, but most of them are based on estimating and subtracting a scatter sinogram from the measured data. We have previously shown that both true and scattered coincidences can be treated similarly by using Compton scattering kinematics to define a locus of scattering which may in turn be used to reconstruct the activity distribution using a generalized scatter maximum-likelihood expectation maximization (GS-MLEM) algorithm. The annihilation position can be further confined by taking advantage of the patient outline (or a geometrical shape that encompasses the patient outline). The proposed method was tested on a phantom generated using GATE. The results have shown that for scatter fractions of 10–60% this algorithm improves the contrast recovery coefficients (CRC) by 4 to 28.6% for a source and 5.1 to 40% for a cold source while the relative standard deviation (RSD) was reduced. Including scattered photons directly into the reconstruction eliminates the need for (often empirical) scatter corrections, and further improvements in the contrast and noise properties of the reconstructed images can be made by including the patient outline in the reconstruction algorithm as a constraint.
- Is Part Of:
- ISRN biomedical imaging. Volume 2013(2013)
- Journal:
- ISRN biomedical imaging
- Issue:
- Volume 2013(2013)
- Issue Display:
- Volume 2013, Issue 2013 (2013)
- Year:
- 2013
- Volume:
- 2013
- Issue:
- 2013
- Issue Sort Value:
- 2013-2013-2013-0000
- Page Start:
- Page End:
- Publication Date:
- 2013-05-02
- Subjects:
- Diagnostic imaging Periodicals
616.0754 - Journal URLs:
- https://www.hindawi.com/journals/isrn/contents/isrn.biomedical.imaging/ ↗
- DOI:
- 10.1155/2013/326847 ↗
- Languages:
- English
- ISSNs:
- 2314-5412
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library HMNTS - ELD Digital store
- Ingest File: