Optimum size of a calibration phantom for x-ray CT to convert the Hounsfield units to stopping power ratios in charged particle therapy treatment planning. (31st October 2017)
- Record Type:
- Journal Article
- Title:
- Optimum size of a calibration phantom for x-ray CT to convert the Hounsfield units to stopping power ratios in charged particle therapy treatment planning. (31st October 2017)
- Main Title:
- Optimum size of a calibration phantom for x-ray CT to convert the Hounsfield units to stopping power ratios in charged particle therapy treatment planning
- Authors:
- Inaniwa, T
Tashima, H
Kanematsu, N - Abstract:
- Abstract: In charged-particle therapy treatment planning, the volumetric distribution of stopping power ratios (SPRs) of body tissues relative to water is used for patient dose calculation. The distribution is conventionally obtained from computed tomography (CT) images of a patient using predetermined conversion functions from the CT numbers to the SPRs. One of the biggest uncertainty sources of patient SPR estimation is insufficient correction of beam hardening arising from the mismatch between the size of the patient cross section and the calibration phantom for producing the conversion functions. The uncertainty would be minimized by selecting a suitable size for the cylindrical water calibration phantom, referred to as an 'effective size' of the patient cross section, L effective . We investigated the L effective for pelvis, abdomen, thorax, and head and neck regions by simulating an ideal CT system using volumetric models of the reference male and female phantoms. The L effective values were 23.3, 20.3, 22.7 and 18.8 cm for the pelvis, abdomen, thorax, and head and neck regions, respectively, and the L effective for whole body was 21.0 cm. Using the conversion function for a 21.0-cm-diameter cylindrical water phantom, we could reduce the root mean square deviation of the SPRs and their mean deviation to ≤0.011 and ≤0.001, respectively, in the whole body. Accordingly, for simplicity, the effective size of 21.0 cm can be used for the whole body, irrespective of body-partAbstract: In charged-particle therapy treatment planning, the volumetric distribution of stopping power ratios (SPRs) of body tissues relative to water is used for patient dose calculation. The distribution is conventionally obtained from computed tomography (CT) images of a patient using predetermined conversion functions from the CT numbers to the SPRs. One of the biggest uncertainty sources of patient SPR estimation is insufficient correction of beam hardening arising from the mismatch between the size of the patient cross section and the calibration phantom for producing the conversion functions. The uncertainty would be minimized by selecting a suitable size for the cylindrical water calibration phantom, referred to as an 'effective size' of the patient cross section, L effective . We investigated the L effective for pelvis, abdomen, thorax, and head and neck regions by simulating an ideal CT system using volumetric models of the reference male and female phantoms. The L effective values were 23.3, 20.3, 22.7 and 18.8 cm for the pelvis, abdomen, thorax, and head and neck regions, respectively, and the L effective for whole body was 21.0 cm. Using the conversion function for a 21.0-cm-diameter cylindrical water phantom, we could reduce the root mean square deviation of the SPRs and their mean deviation to ≤0.011 and ≤0.001, respectively, in the whole body. Accordingly, for simplicity, the effective size of 21.0 cm can be used for the whole body, irrespective of body-part regions for treatment planning in clinical practice. … (more)
- Is Part Of:
- Journal of radiation research. Volume 59:Number 2(2018:Mar.)
- Journal:
- Journal of radiation research
- Issue:
- Volume 59:Number 2(2018:Mar.)
- Issue Display:
- Volume 59, Issue 2 (2018)
- Year:
- 2018
- Volume:
- 59
- Issue:
- 2
- Issue Sort Value:
- 2018-0059-0002-0000
- Page Start:
- 216
- Page End:
- 224
- Publication Date:
- 2017-10-31
- Subjects:
- beam hardening effect -- computed tomography simulation -- standard body tissues -- computational human phantom -- charged particle therapy -- treatment planning
Radiology, Medical -- Periodicals
Radiobiology -- Periodicals
Radiation -- Periodicals
616.0757 - Journal URLs:
- http://bibpurl.oclc.org/web/15847 ↗
http://bibpurl.oclc.org/web/7828 ↗
http://www.journalarchive.jst.go.jp/english/jnltop_en.php?cdjournal=jrr1960 ↗
https://www.jstage.jst.go.jp/browse/jrr ↗
http://ukcatalogue.oup.com/ ↗ - DOI:
- 10.1093/jrr/rrx059 ↗
- Languages:
- English
- ISSNs:
- 0449-3060
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 12364.xml