Implementing safe and robust Total Marrow Irradiation using Helical Tomotherapy – A practical guide. (April 2019)
- Record Type:
- Journal Article
- Title:
- Implementing safe and robust Total Marrow Irradiation using Helical Tomotherapy – A practical guide. (April 2019)
- Main Title:
- Implementing safe and robust Total Marrow Irradiation using Helical Tomotherapy – A practical guide
- Authors:
- Haraldsson, André
Engellau, Jacob
Lenhoff, Stig
Engelholm, Silke
Bäck, Sven
Engström, Per E. - Abstract:
- Highlights: The average dose to key OARs was reduced by up to 40% compared to TBI. Probability of dose coverage to target of 98% dose to 93% volume was 90% for all fractions. Average treatment time was 1 h 56 min, reduced to 1 h 31 min with surface scanning. Total body surface scanning can reduce total time down to 1 h. Abstract: Total Marrow Irradiation (TMI) with Helical Tomotherapy is a radiotherapy treatment technique that targets bone marrow and sanctuary sites prior to stem cell or bone marrow transplantation (SCT/BMT). TMI is a complex procedure that involves several critical steps that all need to be carefully addressed for a successful implementation, such as dose homogeneity in field junctions, choice of target margins, integrity of treatment and back-up planning. In this work we present our solution for a robust and reproducible workflow throughout the treatment chain and data for twenty-three patients treated to date. Material & Methods: Patients were immobilized in a whole body vacuum cushion and thermoplastic mask. CT-scanning and treatment were performed in two parts with field matching at the upper thigh. Target consisted of marrow containing bone and sanctuary sites. Lungs, kidneys, bowel, heart and liver were defined as organs at risk (OAR). A fast surface scanning system was used to position parts of the body not covered by the imaging system (MVCT) as well as to reduce treatment time. Results: All patients completed their treatment and could proceed withHighlights: The average dose to key OARs was reduced by up to 40% compared to TBI. Probability of dose coverage to target of 98% dose to 93% volume was 90% for all fractions. Average treatment time was 1 h 56 min, reduced to 1 h 31 min with surface scanning. Total body surface scanning can reduce total time down to 1 h. Abstract: Total Marrow Irradiation (TMI) with Helical Tomotherapy is a radiotherapy treatment technique that targets bone marrow and sanctuary sites prior to stem cell or bone marrow transplantation (SCT/BMT). TMI is a complex procedure that involves several critical steps that all need to be carefully addressed for a successful implementation, such as dose homogeneity in field junctions, choice of target margins, integrity of treatment and back-up planning. In this work we present our solution for a robust and reproducible workflow throughout the treatment chain and data for twenty-three patients treated to date. Material & Methods: Patients were immobilized in a whole body vacuum cushion and thermoplastic mask. CT-scanning and treatment were performed in two parts with field matching at the upper thigh. Target consisted of marrow containing bone and sanctuary sites. Lungs, kidneys, bowel, heart and liver were defined as organs at risk (OAR). A fast surface scanning system was used to position parts of the body not covered by the imaging system (MVCT) as well as to reduce treatment time. Results: All patients completed their treatment and could proceed with SCT/BMT. Doses to OARs were significantly reduced and target dose homogeneity was improved compared to TBI. Robustness tests performed on field matching and patient positioning support that the field junction technique is adequate. Replacing MVCT with optical surface scanning reduced the treatment time by 25 min per fraction. Conclusion: The methodology presented here has shown to provide a safe, robust and reproducible treatment for Total Marrow Irradiation using Tomotherapy. … (more)
- Is Part Of:
- Physica medica. Volume 60(2019)
- Journal:
- Physica medica
- Issue:
- Volume 60(2019)
- Issue Display:
- Volume 60, Issue 2019 (2019)
- Year:
- 2019
- Volume:
- 60
- Issue:
- 2019
- Issue Sort Value:
- 2019-0060-2019-0000
- Page Start:
- 162
- Page End:
- 167
- Publication Date:
- 2019-04
- Subjects:
- TMI -- Total -- Marrow -- Irradiation -- Radiotherapy -- Tomotherapy -- Helical
Medical physics -- Periodicals
Biophysics -- Periodicals
Biophysics -- Periodicals
Imagerie médicale -- Périodiques
Radiothérapie -- Périodiques
Rayons X -- Sécurité -- Mesures -- Périodiques
Physique -- Périodiques
Médecine -- Périodiques
610.153 - Journal URLs:
- http://www.sciencedirect.com/science/journal/11201797 ↗
http://www.clinicalkey.com/dura/browse/journalIssue/11201797 ↗
http://www.clinicalkey.com.au/dura/browse/journalIssue/11201797 ↗
http://www.elsevier.com/journals ↗
http://www.physicamedica.com ↗ - DOI:
- 10.1016/j.ejmp.2019.03.032 ↗
- Languages:
- English
- ISSNs:
- 1120-1797
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 6475.070000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 12290.xml