Voltage and pace-capture mapping of linear ablation lesions overestimates chronic ablation gap size. Issue 12 (26th April 2018)
- Record Type:
- Journal Article
- Title:
- Voltage and pace-capture mapping of linear ablation lesions overestimates chronic ablation gap size. Issue 12 (26th April 2018)
- Main Title:
- Voltage and pace-capture mapping of linear ablation lesions overestimates chronic ablation gap size
- Authors:
- O'Neill, Louisa
Harrison, James
Chubb, Henry
Whitaker, John
Mukherjee, Rahul K
Bloch, Lars Ølgaard
Andersen, Niels Peter
Dam, Høgni
Jensen, Henrik K
Niederer, Steven
Wright, Matthew
O'Neill, Mark
Williams, Steven E - Abstract:
- Abstract: Aims: Conducting gaps in lesion sets are a major reason for failure of ablation procedures. Voltage mapping and pace-capture have been proposed for intra-procedural identification of gaps. We aimed to compare gap size measured acutely and chronically post-ablation to macroscopic gap size in a porcine model. Methods and results: Intercaval linear ablation was performed in eight Göttingen minipigs with a deliberate gap of ∼5 mm left in the ablation line. Gap size was measured by interpolating ablation contact force values between ablation tags and thresholding at a low force cut-off of 5 g. Bipolar voltage mapping and pace-capture mapping along the length of the line were performed immediately, and at 2 months, post-ablation. Animals were euthanized and gap sizes were measured macroscopically. Voltage thresholds to define scar were determined by receiver operating characteristic analysis as <0.56 mV (acutely) and <0.62 mV (chronically). Taking the macroscopic gap size as gold standard, error in gap measurements were determined for voltage, pace-capture, and ablation contact force maps. All modalities overestimated chronic gap size, by 1.4 ± 2.0 mm (ablation contact force map), 5.1 ± 3.4 mm (pace-capture), and 9.5 ± 3.8 mm (voltage mapping). Error on ablation contact force map gap measurements were significantly less than for voltage mapping ( P = 0.003, Tukey's multiple comparisons test). Chronically, voltage mapping and pace-capture mapping overestimatedAbstract: Aims: Conducting gaps in lesion sets are a major reason for failure of ablation procedures. Voltage mapping and pace-capture have been proposed for intra-procedural identification of gaps. We aimed to compare gap size measured acutely and chronically post-ablation to macroscopic gap size in a porcine model. Methods and results: Intercaval linear ablation was performed in eight Göttingen minipigs with a deliberate gap of ∼5 mm left in the ablation line. Gap size was measured by interpolating ablation contact force values between ablation tags and thresholding at a low force cut-off of 5 g. Bipolar voltage mapping and pace-capture mapping along the length of the line were performed immediately, and at 2 months, post-ablation. Animals were euthanized and gap sizes were measured macroscopically. Voltage thresholds to define scar were determined by receiver operating characteristic analysis as <0.56 mV (acutely) and <0.62 mV (chronically). Taking the macroscopic gap size as gold standard, error in gap measurements were determined for voltage, pace-capture, and ablation contact force maps. All modalities overestimated chronic gap size, by 1.4 ± 2.0 mm (ablation contact force map), 5.1 ± 3.4 mm (pace-capture), and 9.5 ± 3.8 mm (voltage mapping). Error on ablation contact force map gap measurements were significantly less than for voltage mapping ( P = 0.003, Tukey's multiple comparisons test). Chronically, voltage mapping and pace-capture mapping overestimated macroscopic gap size by 11.9 ± 3.7 and 9.8 ± 3.5 mm, respectively. Conclusion: Bipolar voltage and pace-capture mapping overestimate the size of chronic gap formation in linear ablation lesions. The most accurate estimation of chronic gap size was achieved by analysis of catheter-myocardium contact force during ablation. … (more)
- Is Part Of:
- Europace. Volume 20:Issue 12(2018)
- Journal:
- Europace
- Issue:
- Volume 20:Issue 12(2018)
- Issue Display:
- Volume 20, Issue 12 (2018)
- Year:
- 2018
- Volume:
- 20
- Issue:
- 12
- Issue Sort Value:
- 2018-0020-0012-0000
- Page Start:
- 2028
- Page End:
- 2035
- Publication Date:
- 2018-04-26
- Subjects:
- Ablation gaps -- Voltage mapping -- Pace-capture mapping
Arrhythmia -- Treatment -- Periodicals
Cardiac pacing -- Periodicals
Catheter ablation -- Periodicals
Heart -- Physiology -- Periodicals
Electrophysiology -- Periodicals
617.4120645 - Journal URLs:
- http://europace.oxfordjournals.org/ ↗
http://ukcatalogue.oup.com/ ↗ - DOI:
- 10.1093/europace/euy062 ↗
- Languages:
- English
- ISSNs:
- 1099-5129
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 3829.340450
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 12206.xml