Midulla, Fabio - Abstract:
- Abstract: OBJECTIVES: We describe the way we treated 7 children with critical long-term complications after metallic balloon-expandable stenting in the left mainstem bronchus. METHODS: Endoscopic follow-up included a first bronchoscopy 3 weeks after stenting, then monthly for 3 months, every 4–6 months up to 1 year and at scheduled times to calibrate stent diameter up to final calibration. When major complications occurred, patients underwent chest computed tomographic angiography. RESULTS: In 1 of the 7 children (median age 2.8 years), metallic left bronchial stenting served as a bridge to surgery. After a median 4-year follow-up, all 7 children experienced recurrent stent ovalizations with stent breakage in 3 and erosion in 1. In 4 children, computed tomographic angiography showed abundant peribronchial fibrous tissue, in 2 left mediastinal rotation and in 1 displacement along the left bronchus after pulmonary re-expansion as the cause of stent-related complication. Of the 7 children, 6 underwent surgery (5 posterior aortopexy and 1 section of the ligamentum arteriosus) and 3 required nitinol stents placement within the metallic ones. One patient completed the follow-up, and 1 patient was lost to follow-up. All 5 remaining children still have permanent bronchial stents in place, patent and re-epithelialized after a median 10.5-year follow-up. There were no deaths. CONCLUSIONS: Satisfactory anatomical relationships when children have stents placed in the left mainstemAbstract: OBJECTIVES: We describe the way we treated 7 children with critical long-term complications after metallic balloon-expandable stenting in the left mainstem bronchus. METHODS: Endoscopic follow-up included a first bronchoscopy 3 weeks after stenting, then monthly for 3 months, every 4–6 months up to 1 year and at scheduled times to calibrate stent diameter up to final calibration. When major complications occurred, patients underwent chest computed tomographic angiography. RESULTS: In 1 of the 7 children (median age 2.8 years), metallic left bronchial stenting served as a bridge to surgery. After a median 4-year follow-up, all 7 children experienced recurrent stent ovalizations with stent breakage in 3 and erosion in 1. In 4 children, computed tomographic angiography showed abundant peribronchial fibrous tissue, in 2 left mediastinal rotation and in 1 displacement along the left bronchus after pulmonary re-expansion as the cause of stent-related complication. Of the 7 children, 6 underwent surgery (5 posterior aortopexy and 1 section of the ligamentum arteriosus) and 3 required nitinol stents placement within the metallic ones. One patient completed the follow-up, and 1 patient was lost to follow-up. All 5 remaining children still have permanent bronchial stents in place, patent and re-epithelialized after a median 10.5-year follow-up. There were no deaths. CONCLUSIONS: Satisfactory anatomical relationships when children have stents placed in the left mainstem bronchus alone do not guarantee the final success. Several mechanisms intervene to cause critical stent-related complications in children during growth. Permanent metallic stents should be used carefully, and only in selected patients. … (more)
- Is Part Of:
- European journal of cardio-thoracic surgery. Volume 53:Number 3(2018)
- Journal:
- European journal of cardio-thoracic surgery
- Issue:
- Volume 53:Number 3(2018)
- Issue Display:
- Volume 53, Issue 3 (2018)
- Year:
- 2018
- Volume:
- 53
- Issue:
- 3
- Issue Sort Value:
- 2018-0053-0003-0000
- Page Start:
- 610
- Page End:
- 617
- Publication Date:
- 2017-11-08
- Subjects:
- Airway stenting -- Left mainstem bronchus -- Children -- Complication
Heart -- Surgery -- Periodicals
Chest -- Surgery -- Periodicals
617.54 - Journal URLs:
- http://ejcts.oxfordjournals.org/ ↗
http://www.sciencedirect.com/science/journal/10107940 ↗
http://ukcatalogue.oup.com/ ↗ - DOI:
- 10.1093/ejcts/ezx374 ↗
- Languages:
- English
- ISSNs:
- 1010-7940
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 3829.725620
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 12205.xml