The unstable pacing thresholds of the leadless transcatheter pacemaker affected by body positions in subacute phase after implant. Issue 1 (27th December 2018)
- Record Type:
- Journal Article
- Title:
- The unstable pacing thresholds of the leadless transcatheter pacemaker affected by body positions in subacute phase after implant. Issue 1 (27th December 2018)
- Main Title:
- The unstable pacing thresholds of the leadless transcatheter pacemaker affected by body positions in subacute phase after implant
- Authors:
- Yoh, Masue
Takagi, Masahiko
Takahashi, Hiroki
Yoshio, Takuro
Shiojima, Ichiro - Editors:
- Dinov, Borislav
Schönbauer, Robert
Yousef Abumuaileq, Rami Riziq
Akhtar, Mohammed
Patel, Peysh A - Abstract:
- Abstract: Background: If the threshold at implant of leadless transcatheter pacemakers (LTPs) is less than 2.0 V, pacing thresholds reportedly decrease significantly by 1 month and maintain an optimal value of less than 1.0 V by 6 months. Case summary: We report a case series of two patients with unstable pacing thresholds of the LTPs in the subacute phase after implant. The first patient (77-year-old man) was implanted an LTP for sick sinus syndrome. At that time of implant, the pacing threshold was 0.9 V at 0.24 ms. At 1 week and 1 month later, the threshold had increased to more than 2.0 V at 0.24 ms. We investigated the trend data for the week and found variations in the threshold. The second patient (81-year-old man) was implanted an LTP for bradycardia and atrial fibrillation. The pacing threshold at implantation was 0.63 V at 0.24 ms. One week later, the threshold had increased in supine position and decreased in sitting position. The trend data for the week were fluctuating greatly. Discussion: The pacing threshold may increase to more than 2.0 V with significant fluctuation on assessment at 1 week and 1 month after implantation in association with changes in body position, even though we confirmed a stable threshold at implant. If an increased threshold is observed, it is necessary to check the trend data and threshold in each body position.
- Is Part Of:
- European heart journal. Volume 3:Issue 1(2019)
- Journal:
- European heart journal
- Issue:
- Volume 3:Issue 1(2019)
- Issue Display:
- Volume 3, Issue 1 (2019)
- Year:
- 2019
- Volume:
- 3
- Issue:
- 1
- Issue Sort Value:
- 2019-0003-0001-0000
- Page Start:
- Page End:
- Publication Date:
- 2018-12-27
- Subjects:
- Leadless pacemaker -- Threshold Body position -- Subacute phase -- Fluctuation -- Case series
Cardiology -- Periodicals
Cardiology -- Case studies -- Periodicals
Heart -- Diseases -- Periodicals
Heart -- Diseases -- Case studies -- Periodicals
616.12 - Journal URLs:
- http://www.oxfordjournals.org/ ↗
https://academic.oup.com/ehjcr/ ↗ - DOI:
- 10.1093/ehjcr/yty160 ↗
- Languages:
- English
- ISSNs:
- 2514-2119
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 11993.xml