Commissioning of the world's first compact pencil‐beam scanning proton therapy system. (20th November 2017)
- Record Type:
- Journal Article
- Title:
- Commissioning of the world's first compact pencil‐beam scanning proton therapy system. (20th November 2017)
- Main Title:
- Commissioning of the world's first compact pencil‐beam scanning proton therapy system
- Authors:
- Pidikiti, Rajesh
Patel, Bijal C.
Maynard, Matthew R.
Dugas, Joseph P.
Syh, Joseph
Sahoo, Narayan
Wu, Hsinshun Terry
Rosen, Lane R. - Abstract:
- Abstract: This paper summarizes clinical commissioning of the world's first commercial, clinically utilized installation of a compact, image‐guided, pencil‐beam scanning, intensity‐modulated proton therapy system, the IBA Proteus ® ONE, at the Willis‐Knighton Cancer Center (WKCC) in Shreveport, LA. The Proteus ® ONE is a single‐room, compact‐gantry system employing a cyclotron‐generated proton beam with image guidance via cone‐beam CT as well as stereoscopic orthogonal and oblique planar kV imaging. Coupling 220° of gantry rotation with a 6D robotic couch capable of in plane patient rotations of over 180° degrees allows for 360° of treatment access. Along with general machine characterization, system commissioning required: (a) characterization and calibration of the proton beam, (b) treatment planning system commissioning including CT‐to‐density curve determination, (c) image guidance system commissioning, and (d) safety verification (interlocks and radiation survey). System readiness for patient treatment was validated by irradiating calibration TLDs as well as prostate, head, and lung phantoms from the Imaging and Radiation Oncology Core (IROC), Houston. These results confirmed safe and accurate machine functionality suitable for patient treatment. WKCC also successfully completed an on‐site dosimetry review by an independent team of IROC physicists that corroborated accurate Proteus ® ONE dosimetry.
- Is Part Of:
- Journal of applied clinical medical physics. Volume 19:Number 1(2018)
- Journal:
- Journal of applied clinical medical physics
- Issue:
- Volume 19:Number 1(2018)
- Issue Display:
- Volume 19, Issue 1 (2018)
- Year:
- 2018
- Volume:
- 19
- Issue:
- 1
- Issue Sort Value:
- 2018-0019-0001-0000
- Page Start:
- 94
- Page End:
- 105
- Publication Date:
- 2017-11-20
- Subjects:
- clinical commissioning -- PBS -- pencil beam -- Proteus®ONE -- proton -- radiotherapy -- spot scanning
Medical physics -- Periodicals
Clinical medicine -- Periodicals
Health Physics
Clinical Medicine
Electronic journals
Periodicals
Periodicals
Fulltext
Internet Resources
610.153 - Journal URLs:
- http://aapm.onlinelibrary.wiley.com/hub/journal/10.1002/(ISSN)1526-9914/ ↗
http://bibpurl.oclc.org/web/7294 ↗
http://www.jacmp.org/ ↗
http://onlinelibrary.wiley.com/ ↗ - DOI:
- 10.1002/acm2.12225 ↗
- Languages:
- English
- ISSNs:
- 1526-9914
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 11954.xml