Comparison of the Convolution algorithm with TMR10 for Leksell Gamma knife and dosimetric verification with radiochromic gel dosimeter. (10th December 2017)
- Record Type:
- Journal Article
- Title:
- Comparison of the Convolution algorithm with TMR10 for Leksell Gamma knife and dosimetric verification with radiochromic gel dosimeter. (10th December 2017)
- Main Title:
- Comparison of the Convolution algorithm with TMR10 for Leksell Gamma knife and dosimetric verification with radiochromic gel dosimeter
- Authors:
- Osmancikova, Petra
Novotny, Josef
Solc, Jaroslav
Pipek, Jan - Abstract:
- Abstract: The Convolution algorithm, implemented in Leksell GammaPlan ® ver. Here, 10, is the first algorithm for Leksell Gamma Knife that takes heterogeneities into account and models dose build‐up effects close to tissue boundaries. The aim of this study was preliminary comparison of the Convolution and TMR10 algorithms for real clinical cases and dosimetric verification of the algorithms, using measurements in a phantom. A total of 25 patients involved in comparison of the Convolution and TMR10 algorithms were divided into three groups: patients with benign tumors close to heterogeneities, patients with functional disorders, and patients with tumors located far from heterogeneities. Differences were observed especially in the group of patients with tumors close to heterogeneities, where the difference in maximal dose to critical structures for the Convolution algorithm was up to 15% compared to the TMR10 algorithm. Dosimetric verification of the algorithm was performed, using a radiochromic gel dosimeter based on Turnbull blue dye in a special heterogeneous phantom. Relative dose distributions measured with the radiochromic gel dosimeter agreed very well with both the TMR10 and Convolution calculations. We observed small discrepancies in the direction in which the largest inhomogeneity was positioned. Verification results indicated that the Convolution algorithm provides a different dose distribution, especially in regions close to heterogeneities and particularly forAbstract: The Convolution algorithm, implemented in Leksell GammaPlan ® ver. Here, 10, is the first algorithm for Leksell Gamma Knife that takes heterogeneities into account and models dose build‐up effects close to tissue boundaries. The aim of this study was preliminary comparison of the Convolution and TMR10 algorithms for real clinical cases and dosimetric verification of the algorithms, using measurements in a phantom. A total of 25 patients involved in comparison of the Convolution and TMR10 algorithms were divided into three groups: patients with benign tumors close to heterogeneities, patients with functional disorders, and patients with tumors located far from heterogeneities. Differences were observed especially in the group of patients with tumors close to heterogeneities, where the difference in maximal dose to critical structures for the Convolution algorithm was up to 15% compared to the TMR10 algorithm. Dosimetric verification of the algorithm was performed, using a radiochromic gel dosimeter based on Turnbull blue dye in a special heterogeneous phantom. Relative dose distributions measured with the radiochromic gel dosimeter agreed very well with both the TMR10 and Convolution calculations. We observed small discrepancies in the direction in which the largest inhomogeneity was positioned. Verification results indicated that the Convolution algorithm provides a different dose distribution, especially in regions close to heterogeneities and particularly for lower isodose volumes. However, the results obtained with gamma analyses in the gel dosimetry experiment did not verify the assumption that the Convolution algorithm provides more accurate dose calculation. … (more)
- Is Part Of:
- Journal of applied clinical medical physics. Volume 19:Number 1(2018)
- Journal:
- Journal of applied clinical medical physics
- Issue:
- Volume 19:Number 1(2018)
- Issue Display:
- Volume 19, Issue 1 (2018)
- Year:
- 2018
- Volume:
- 19
- Issue:
- 1
- Issue Sort Value:
- 2018-0019-0001-0000
- Page Start:
- 138
- Page End:
- 144
- Publication Date:
- 2017-12-10
- Subjects:
- gel dosimetry -- stereotactic radiosurgery -- verification
Medical physics -- Periodicals
Clinical medicine -- Periodicals
Health Physics
Clinical Medicine
Electronic journals
Periodicals
Periodicals
Fulltext
Internet Resources
610.153 - Journal URLs:
- http://aapm.onlinelibrary.wiley.com/hub/journal/10.1002/(ISSN)1526-9914/ ↗
http://bibpurl.oclc.org/web/7294 ↗
http://www.jacmp.org/ ↗
http://onlinelibrary.wiley.com/ ↗ - DOI:
- 10.1002/acm2.12238 ↗
- Languages:
- English
- ISSNs:
- 1526-9914
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 11954.xml