Validation of the RayStation Monte Carlo dose calculation algorithm using realistic animal tissue phantoms. Issue 10 (21st September 2019)
- Record Type:
- Journal Article
- Title:
- Validation of the RayStation Monte Carlo dose calculation algorithm using realistic animal tissue phantoms. Issue 10 (21st September 2019)
- Main Title:
- Validation of the RayStation Monte Carlo dose calculation algorithm using realistic animal tissue phantoms
- Authors:
- Schreuder, Andries N.
Bridges, Daniel S.
Rigsby, Lauren
Blakey, Marc
Janson, Martin
Hedrick, Samantha G.
Wilkinson, John B. - Abstract:
- Abstract: Purpose: The aim of this study is to validate the RayStation Monte Carlo (MC) dose algorithm using animal tissue neck phantoms and a water breast phantom. Methods: Three anthropomorphic phantoms were used in a clinical setting to test the RayStation MC dose algorithm. We used two real animal necks that were cut to a workable shape while frozen and then thawed before being CT scanned. Secondly, we made a patient breast phantom using a breast prosthesis filled with water and placed on a flat surface. Dose distributions in the animal and breast phantoms were measured using the MatriXX PT device. Results: The measured doses to the neck and breast phantoms compared exceptionally well with doses calculated by the analytical pencil beam (APB) and MC algorithms. The comparisons between APB and MC dose calculations and MatriXX PT measurements yielded an average depth difference for best gamma agreement of <1 mm for the neck phantoms. For the breast phantom better average gamma pass rates between measured and calculated dose distributions were observed for the MC than for the APB algorithms. Conclusions: The MC dose calculations are more accurate than the APB calculations for the static phantoms conditions we evaluated, especially in areas where significant inhomogeneous interfaces are traversed by the beam.
- Is Part Of:
- Journal of applied clinical medical physics. Volume 20:Issue 10(2019)
- Journal:
- Journal of applied clinical medical physics
- Issue:
- Volume 20:Issue 10(2019)
- Issue Display:
- Volume 20, Issue 10 (2019)
- Year:
- 2019
- Volume:
- 20
- Issue:
- 10
- Issue Sort Value:
- 2019-0020-0010-0000
- Page Start:
- 160
- Page End:
- 171
- Publication Date:
- 2019-09-21
- Subjects:
- analytical dose algorithms -- Monte Carlo -- pencil beam scanning -- proton therapy -- spot scanning
Medical physics -- Periodicals
Clinical medicine -- Periodicals
Health Physics
Clinical Medicine
Electronic journals
Periodicals
Periodicals
Fulltext
Internet Resources
610.153 - Journal URLs:
- http://aapm.onlinelibrary.wiley.com/hub/journal/10.1002/(ISSN)1526-9914/ ↗
http://bibpurl.oclc.org/web/7294 ↗
http://www.jacmp.org/ ↗
http://onlinelibrary.wiley.com/ ↗ - DOI:
- 10.1002/acm2.12733 ↗
- Languages:
- English
- ISSNs:
- 1526-9914
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 11894.xml