Long-term results after the Ross procedure with the decellularized AutoTissue Matrix P® bioprosthesis used for pulmonary valve replacement. (1st December 2018)
- Record Type:
- Journal Article
- Title:
- Long-term results after the Ross procedure with the decellularized AutoTissue Matrix P® bioprosthesis used for pulmonary valve replacement. (1st December 2018)
- Main Title:
- Long-term results after the Ross procedure with the decellularized AutoTissue Matrix P® bioprosthesis used for pulmonary valve replacement
- Authors:
- Christ, Torsten
Paun, Alexandru Claudiu
Grubitzsch, Herko
Holinski, Sebastian
Falk, Volkmar
Dushe, Simon - Abstract:
- Abstract: OBJECTIVES: Since 1967, the Ross procedure has been performed to treat aortic valve disease using homografts for pulmonary valve replacement. The decellularized Matrix P ® prosthesis was developed to overcome (some) limitations of homografts. Until now, the long-term outcome data have been unavailable. METHODS: Between 2002 and 2010, the Ross procedures using the Matrix P prosthesis were performed in 492 adult patients (mean age 57.2 ± 10.6 years, range 21–73 years) at our institution. Patient data were prospectively collected and analysed (3617.3 patient-years, mean follow-up 7.7 ± 4.3 years). Completeness of follow-up at 1, 5 and 10 years was 98.4%, 94.5% and 91.0%, respectively. RESULTS: Hospital mortality was 3.9% ( n = 19). During follow-up, 121 patients died resulting in a survival rate at 5, 10 and 12.5 years of 82.8 ± 1.7%, 70.4 ± 2.3% and 62.4 ± 2.9%, respectively. Echocardiography revealed a high incidence of relevant dysfunction of the Matrix P prosthesis and subsequent right ventricular failure. Primary reoperation/reintervention was necessary for 150 Matrix P and 48 autografts. Freedom from pulmonary valve reoperation at 5, 10 and 12.5 years was 76.2 ± 2.1%, 58.6 ± 2.9% and 53.4 ± 3.4%, respectively. The autograft function and the left ventricular function showed similar results as previously reported with a freedom from autograft reoperation at 5, 10 and 12.5 years of 91.8 ± 1.4%, 86.1 ± 2.0% and 86.1 ± 2.0%, respectively.Abstract: OBJECTIVES: Since 1967, the Ross procedure has been performed to treat aortic valve disease using homografts for pulmonary valve replacement. The decellularized Matrix P ® prosthesis was developed to overcome (some) limitations of homografts. Until now, the long-term outcome data have been unavailable. METHODS: Between 2002 and 2010, the Ross procedures using the Matrix P prosthesis were performed in 492 adult patients (mean age 57.2 ± 10.6 years, range 21–73 years) at our institution. Patient data were prospectively collected and analysed (3617.3 patient-years, mean follow-up 7.7 ± 4.3 years). Completeness of follow-up at 1, 5 and 10 years was 98.4%, 94.5% and 91.0%, respectively. RESULTS: Hospital mortality was 3.9% ( n = 19). During follow-up, 121 patients died resulting in a survival rate at 5, 10 and 12.5 years of 82.8 ± 1.7%, 70.4 ± 2.3% and 62.4 ± 2.9%, respectively. Echocardiography revealed a high incidence of relevant dysfunction of the Matrix P prosthesis and subsequent right ventricular failure. Primary reoperation/reintervention was necessary for 150 Matrix P and 48 autografts. Freedom from pulmonary valve reoperation at 5, 10 and 12.5 years was 76.2 ± 2.1%, 58.6 ± 2.9% and 53.4 ± 3.4%, respectively. The autograft function and the left ventricular function showed similar results as previously reported with a freedom from autograft reoperation at 5, 10 and 12.5 years of 91.8 ± 1.4%, 86.1 ± 2.0% and 86.1 ± 2.0%, respectively. CONCLUSIONS: The Matrix P prosthesis used for the right ventricular outflow tract reconstruction in the Ross procedure showed unfavourable long-term echocardiographic results with a high rate of reoperation/reintervention for structural pulmonary valve failure. As a consequence, long-term survival of this patient cohort was impaired. Based on these findings, the use of the Matrix P prosthesis for pulmonary valve replacement for Ross procedures in adults should not be recommended. … (more)
- Is Part Of:
- European journal of cardio-thoracic surgery. Volume 55:Number 5(2019)
- Journal:
- European journal of cardio-thoracic surgery
- Issue:
- Volume 55:Number 5(2019)
- Issue Display:
- Volume 55, Issue 5 (2019)
- Year:
- 2019
- Volume:
- 55
- Issue:
- 5
- Issue Sort Value:
- 2019-0055-0005-0000
- Page Start:
- 885
- Page End:
- 892
- Publication Date:
- 2018-12-01
- Subjects:
- Ross procedure -- Pulmonary valve replacement -- Aortic valve replacement -- Heart valve prosthesis -- Bioprosthesis
Heart -- Surgery -- Periodicals
Chest -- Surgery -- Periodicals
617.54 - Journal URLs:
- http://ejcts.oxfordjournals.org/ ↗
http://www.sciencedirect.com/science/journal/10107940 ↗
http://ukcatalogue.oup.com/ ↗ - DOI:
- 10.1093/ejcts/ezy377 ↗
- Languages:
- English
- ISSNs:
- 1010-7940
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 3829.725620
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 11794.xml