Minimally invasive percutaneous robotic thoracolumbar pedicle screw implantation combined with three‐dimensional fluoroscopy can reduce radiation: A cadaver and phantom study. Issue 5 (10th July 2019)
- Record Type:
- Journal Article
- Title:
- Minimally invasive percutaneous robotic thoracolumbar pedicle screw implantation combined with three‐dimensional fluoroscopy can reduce radiation: A cadaver and phantom study. Issue 5 (10th July 2019)
- Main Title:
- Minimally invasive percutaneous robotic thoracolumbar pedicle screw implantation combined with three‐dimensional fluoroscopy can reduce radiation: A cadaver and phantom study
- Authors:
- Spyrantis, Andrea
Cattani, Adriano
Seifert, Volker
Freiman, Thomas M.
Setzer, Matthias - Abstract:
- Abstract: Objective: Feasibility of robotic minimal invasive pedicle screw (PS) placement in the thoracolumbar spine using preoperative three‐dimensional (3D) fluoroscopy without computed tomography (CT). Methods: In three cadavers, 12 pedicle screws were implanted in thoracolumbar segments with the robotic surgery assistant. 3D fluoroscopy was performed for preoperative referencing, planning, and identification of postoperative screw position. The radiation exposure of fluoroscopy and a CT scanner was compared, measuring the CT dose index (CTDIw ). Results: Pedicle screw positioning was graded according to the classification of Gertzbein and Robbins: 11 of 12 pedicle screws showed optimal transpedicular position (grade 1) and one was positioned less than 2 mm outside (grade 2). No major deviations were observed. Referencing with 3D fluoroscopy resulted in a CTDIw reduction of 84% in the cervical and 33% in the lumbar spine. Conclusions: Robot‐guided PS placement, using 3D fluoroscopy for referencing, is a reliable tool for minimally invasive PS implantation; radiation exposure can be reduced.
- Is Part Of:
- International journal of medical robotics and computer assisted surgery. Volume 15:Issue 5(2019)
- Journal:
- International journal of medical robotics and computer assisted surgery
- Issue:
- Volume 15:Issue 5(2019)
- Issue Display:
- Volume 15, Issue 5 (2019)
- Year:
- 2019
- Volume:
- 15
- Issue:
- 5
- Issue Sort Value:
- 2019-0015-0005-0000
- Page Start:
- n/a
- Page End:
- n/a
- Publication Date:
- 2019-07-10
- Subjects:
- 3D fluoroscopy -- 3D imaging -- intraoperative imaging -- minimal invasive surgery -- pedicle screw -- preoperative planning -- robotic surgery -- spine
Robotics in medicine -- Periodicals
Surgery -- Technological innovations -- Periodicals
Imaging systems in medicine -- Periodicals
617.90285 - Journal URLs:
- http://onlinelibrary.wiley.com/journal/10.1002/(ISSN)1478-596X ↗
http://www.roboticpublications.com ↗
http://onlinelibrary.wiley.com/ ↗ - DOI:
- 10.1002/rcs.2022 ↗
- Languages:
- English
- ISSNs:
- 1478-5951
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 4542.347800
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 11672.xml