Clinical performance evaluation of the new hematology analyzer Mindray BC‐6000. (9th July 2019)
- Record Type:
- Journal Article
- Title:
- Clinical performance evaluation of the new hematology analyzer Mindray BC‐6000. (9th July 2019)
- Main Title:
- Clinical performance evaluation of the new hematology analyzer Mindray BC‐6000
- Authors:
- Shen, Yimin
Cao, Jun
Zhou, Zhiying
Wang, Yong
Shen, Yili
He, Jun - Abstract:
- Abstract: Introduction: With the progression of blood analysis technology, hematology analyzers become more complex and diverse. How to choose a superb instrument is a challenge for the laboratories. In the essay, we studied whether the newest BC‐6000 hematology analyzer meets the needs of a clinical hematology laboratory. Methods: Methods comparison was performed using 350 blood samples from patients between different measurement procedures; the basic analytical performance was also tested, including the throughput, carryover, precision, and linearity in different modes. The flagging performances for blasts, immature granulocytes, and NRBC were compared with manual microscopy. Results: There were minimal carryover (<0.30%) and excellent actual blood linearity for all routinely used parameters concerned by the clinicians ( R 2 ≥ 0.997). Repeatability and reproducibility were satisfactory at all testing levels. The functional sensitivity of leukocytes and platelets in the blood and leukocytes and erythrocytes in body fluid was excellent at the 20% CV level. BC‐6000 and XN displayed very high correlations for complete blood count (CBC) parameters and very high consistency for leukocyte differentials and NRBC compared with manual microscopy. BC‐6000 showed excellent sensitivity and specificity flagging ability on blasts (82.9% and 82.4%) and NRBC (80.0% and 96.9%). For immature granulocytes, BC‐6000 showed excellent sensitivity but common specificity flagging ability (91.7%Abstract: Introduction: With the progression of blood analysis technology, hematology analyzers become more complex and diverse. How to choose a superb instrument is a challenge for the laboratories. In the essay, we studied whether the newest BC‐6000 hematology analyzer meets the needs of a clinical hematology laboratory. Methods: Methods comparison was performed using 350 blood samples from patients between different measurement procedures; the basic analytical performance was also tested, including the throughput, carryover, precision, and linearity in different modes. The flagging performances for blasts, immature granulocytes, and NRBC were compared with manual microscopy. Results: There were minimal carryover (<0.30%) and excellent actual blood linearity for all routinely used parameters concerned by the clinicians ( R 2 ≥ 0.997). Repeatability and reproducibility were satisfactory at all testing levels. The functional sensitivity of leukocytes and platelets in the blood and leukocytes and erythrocytes in body fluid was excellent at the 20% CV level. BC‐6000 and XN displayed very high correlations for complete blood count (CBC) parameters and very high consistency for leukocyte differentials and NRBC compared with manual microscopy. BC‐6000 showed excellent sensitivity and specificity flagging ability on blasts (82.9% and 82.4%) and NRBC (80.0% and 96.9%). For immature granulocytes, BC‐6000 showed excellent sensitivity but common specificity flagging ability (91.7% and 65.6%). Conclusion: The clinical performance of BC‐6000 is excellent, and the analyzer can provide timely and accurate reporting for most of the small‐ to large‐scale laboratories. … (more)
- Is Part Of:
- International journal of laboratory hematology. Volume 41:Number 5(2019:Oct.)
- Journal:
- International journal of laboratory hematology
- Issue:
- Volume 41:Number 5(2019:Oct.)
- Issue Display:
- Volume 41, Issue 5 (2019)
- Year:
- 2019
- Volume:
- 41
- Issue:
- 5
- Issue Sort Value:
- 2019-0041-0005-0000
- Page Start:
- 622
- Page End:
- 634
- Publication Date:
- 2019-07-09
- Subjects:
- BC‐6000 -- body fluid -- flagging ability -- hematology analyzer -- methods comparison -- performance evaluation -- repeatability -- reproducibility
Hematology -- Periodicals
Blood -- Diseases -- Periodicals
Hematology -- Periodicals
616.15005 - Journal URLs:
- http://firstsearch.oclc.org/FSIP?db=ECO&journal=1751-5521&screen=info&done=referer ↗
http://www.blackwell-synergy.com/loi/clh ↗
http://onlinelibrary.wiley.com/journal/10.1111/(ISSN)1751-553X ↗
http://onlinelibrary.wiley.com/ ↗ - DOI:
- 10.1111/ijlh.13075 ↗
- Languages:
- English
- ISSNs:
- 1751-5521
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 4542.312220
British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 11678.xml