Automated Early Identification of an Excessive Air-in-Oil X-ray Tube Artifact That Mimics Acute Cerebral Infarct. Issue 1 (January 2019)
- Record Type:
- Journal Article
- Title:
- Automated Early Identification of an Excessive Air-in-Oil X-ray Tube Artifact That Mimics Acute Cerebral Infarct. Issue 1 (January 2019)
- Main Title:
- Automated Early Identification of an Excessive Air-in-Oil X-ray Tube Artifact That Mimics Acute Cerebral Infarct
- Authors:
- Jaffe, Tracy A.
Winslow, James
Zhang, Yakun
Allen, Brian C.
Choudhury, Kingshuk R.
Samei, Ehsan - Abstract:
- Abstract : Purpose: This study aimed to develop an automated, quantitative method to increase the likelihood of identifying and preventing such air-in-oil (AIO) artifact. Methods: A 1-dimensional radial representation of the 2-dimensional noise power spectrum (NPS) was calculated from AIO artifact images and compared with artifact-free images. A quality control (QC) software program was modified to include measurements of NPS average frequency within the water section of daily phantom scans. Threshold values for each CT system were incorporated into daily QC. Results: Noise power spectrum for AIO artifact images included a large low-frequency peak compared with artifact-free images; NPS average frequencies were 0.197 and 0.319 line pairs per millimeter for AIO artifact and artifact-free images, respectively. Automated QC successfully identified 3 AIO artifacts before detrimental clinical effect occurred. Conclusions: Serious clinical problems associated with AIO artifact can be detected and avoided by incorporating NPS average frequency measurements of daily phantom images into an automated QC program.
- Is Part Of:
- Journal of computer assisted tomography. Volume 43:Issue 1(2019)
- Journal:
- Journal of computer assisted tomography
- Issue:
- Volume 43:Issue 1(2019)
- Issue Display:
- Volume 43, Issue 1 (2019)
- Year:
- 2019
- Volume:
- 43
- Issue:
- 1
- Issue Sort Value:
- 2019-0043-0001-0000
- Page Start:
- Page End:
- Publication Date:
- 2019-01
- Subjects:
- CT -- artifact -- quality control -- phantom
Tomography -- Periodicals
Tomography -- Periodicals
Tomography
Periodicals
616.0757 - Journal URLs:
- http://journals.lww.com/jcat/pages/default.aspx ↗
http://ovidsp.tx.ovid.com ↗
http://www.jcat.org ↗
http://www.rad.bqsm.edu/jcat ↗
http://journals.lww.com ↗
http://www.lww.com/Product/0363-8715 ↗ - DOI:
- 10.1097/RCT.0000000000000770 ↗
- Languages:
- English
- ISSNs:
- 0363-8715
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 4963.650000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 11511.xml