Cardiac Pulsatility– and Respiratory-Induced Deformations of the Renal Arteries and Snorkel Stents After Snorkel Endovascular Aneurysm Sealing. (August 2019)
- Record Type:
- Journal Article
- Title:
- Cardiac Pulsatility– and Respiratory-Induced Deformations of the Renal Arteries and Snorkel Stents After Snorkel Endovascular Aneurysm Sealing. (August 2019)
- Main Title:
- Cardiac Pulsatility– and Respiratory-Induced Deformations of the Renal Arteries and Snorkel Stents After Snorkel Endovascular Aneurysm Sealing
- Authors:
- Cheng, Christopher P.
Suh, Ga-Young
Kim, John J.
Holden, Andrew - Abstract:
- Purpose: To quantify deformations of renal arteries and snorkel stents after snorkel endovascular aneurysm sealing (Sn-EVAS) resulting from cardiac pulsatility and respiration and compare these deformations to patients with untreated abdominal aortic aneurysms (AAA) and snorkel endovascular aneurysm repair (Sn-EVAR).Materials and Methods: Ten Sn-EVAS patients (mean age 75±6 years; 8 men) were scanned with cardiac-gated, respiration-resolved computed tomography angiography. From 3-dimensional geometric models, changes in renal artery and stent angulation and curvature due to cardiac pulsatility and respiration were quantified. Respiration-induced motions were compared with those of 16 previously reported untreated AAA patients and 11 Sn-EVAR patients.Results: Renal artery bending at the stent end was greater for respiratory vs cardiac influences (6°±7° vs −1°±2°, p<0.025). Respiration caused a 3-fold greater deformation on the left renal artery as compared with the right side. Maximum curvature change was higher for respiratory vs cardiac influences (0.49±0.29 vs 0.24±0.17 cm −1, p<0.025), and snorkel renal stents experienced similar maximum curvature change due to cardiac pulsatility and respiration (0.14±0.10 vs 0.19±0.09 cm −1, p=0.142). When comparing the 3 patient cohorts for respiratory-induced deformation, there was significant renal branch angulation in untreated AAAs, but not in Sn-EVAR or Sn-EVAS, and there was significant bending at the stent end in Sn-EVAR andPurpose: To quantify deformations of renal arteries and snorkel stents after snorkel endovascular aneurysm sealing (Sn-EVAS) resulting from cardiac pulsatility and respiration and compare these deformations to patients with untreated abdominal aortic aneurysms (AAA) and snorkel endovascular aneurysm repair (Sn-EVAR).Materials and Methods: Ten Sn-EVAS patients (mean age 75±6 years; 8 men) were scanned with cardiac-gated, respiration-resolved computed tomography angiography. From 3-dimensional geometric models, changes in renal artery and stent angulation and curvature due to cardiac pulsatility and respiration were quantified. Respiration-induced motions were compared with those of 16 previously reported untreated AAA patients and 11 Sn-EVAR patients.Results: Renal artery bending at the stent end was greater for respiratory vs cardiac influences (6°±7° vs −1°±2°, p<0.025). Respiration caused a 3-fold greater deformation on the left renal artery as compared with the right side. Maximum curvature change was higher for respiratory vs cardiac influences (0.49±0.29 vs 0.24±0.17 cm −1, p<0.025), and snorkel renal stents experienced similar maximum curvature change due to cardiac pulsatility and respiration (0.14±0.10 vs 0.19±0.09 cm −1, p=0.142). When comparing the 3 patient cohorts for respiratory-induced deformation, there was significant renal branch angulation in untreated AAAs, but not in Sn-EVAR or Sn-EVAS, and there was significant bending at the stent end in Sn-EVAR and Sn-EVAS. Maximum curvature change due to respiration was ~10-fold greater in Sn-EVAR and Sn-EVAS compared to untreated AAAs.Conclusion: The findings suggest that cardiac and respiratory influences may challenge the mechanical durability of snorkel stents of Sn-EVAS; similarly, however, respiration may be the primary culprit for tissue irritation, increasing the risk for stent-end thrombosis, especially in the left renal artery. The bending stiffness of snorkel stents in both the Sn-EVAR and Sn-EVAS cohorts damped renal branch angulation while it intensified bending of the artery distal to the snorkel stent. Understanding these device-to-artery interactions is critical as they may affect mechanical durability of branch stents and quality and durability of treatment. … (more)
- Is Part Of:
- Journal of endovascular therapy. Volume 26:Number 4(2019:Aug.)
- Journal:
- Journal of endovascular therapy
- Issue:
- Volume 26:Number 4(2019:Aug.)
- Issue Display:
- Volume 26, Issue 4 (2019)
- Year:
- 2019
- Volume:
- 26
- Issue:
- 4
- Issue Sort Value:
- 2019-0026-0004-0000
- Page Start:
- 556
- Page End:
- 564
- Publication Date:
- 2019-08
- Subjects:
- abdominal aortic aneurysm -- cardiac pulsatility -- chimney stents -- deformation -- endovascular aneurysm repair -- endovascular aneurysm sealing -- mechanical durability -- parallel stents -- renal artery -- respiration -- snorkel stents -- thrombosis
Blood-vessels -- Endoscopic surgery -- Periodicals
Angioscopy -- Periodicals
Intravenous catheterization -- Periodicals
Peripheral vascular diseases -- Treatment -- Periodicals
Vascular Surgical Procedures -- Periodicals
Angioscopy -- Periodicals
Catheterization, Peripheral -- Periodicals
Peripheral Vascular Diseases -- therapy -- Periodicals
Angioscopie
Maladies vasculaires périphériques
617.413 - Journal URLs:
- http://jet.sagepub.com/ ↗
http://www.jevt.org ↗
http://www.uk.sagepub.com ↗ - DOI:
- 10.1177/1526602819856363 ↗
- Languages:
- English
- ISSNs:
- 1526-6028
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 11450.xml