Magnetic Particle Imaging–Guided Stenting. (August 2019)
- Record Type:
- Journal Article
- Title:
- Magnetic Particle Imaging–Guided Stenting. (August 2019)
- Main Title:
- Magnetic Particle Imaging–Guided Stenting
- Authors:
- Herz, Stefan
Vogel, Patrick
Kampf, Thomas
Dietrich, Philipp
Veldhoen, Simon
Rückert, Martin A.
Kickuth, Ralph
Behr, Volker C.
Bley, Thorsten A. - Abstract:
- Purpose: To assess the feasibility of magnetic particle imaging (MPI) to guide stenting in a phantom model.Materials and Methods: MPI is a new tomographic imaging method based on the background-free magnetic field detection of a tracer agent composed of superparamagnetic iron oxide nanoparticles (SPIOs). All experiments were conducted on a custom-built MPI scanner (field of view: 29-mm diameter, 65-mm length; isotropic spatial resolution 1–1.5-mm). Stenosis phantoms (n=3) consisted of polyvinyl chloride (PVC) tubes (8-mm inner diameter) prepared with centrally aligned cable binders to form a ~50% stenosis. A dedicated image reconstruction algorithm allowed precise tracking of endovascular instruments at 8 frames/s with a latency time of ~115 ms. A custom-made MPI-visible lacquer was used to manually label conventional guidewires, balloon catheters, and stainless steel balloon-expandable stents. Vascular stenoses were visualized by injecting a diluted SPIO tracer (ferucarbotran, 10 mmol iron/L) into the vessel phantoms. Balloon angioplasty and stent placement were performed by inflating balloon catheters and stent delivery balloons with diluted ferucarbotran.Results: After deployment of the stent, the markers on its ends were clearly visible. The applied lacquer markers were thin enough to not relevantly alter gliding properties of the devices while withstanding friction during the experiments. Placing an optimized flexible lacquer formulation on the preexisting radiopaquePurpose: To assess the feasibility of magnetic particle imaging (MPI) to guide stenting in a phantom model.Materials and Methods: MPI is a new tomographic imaging method based on the background-free magnetic field detection of a tracer agent composed of superparamagnetic iron oxide nanoparticles (SPIOs). All experiments were conducted on a custom-built MPI scanner (field of view: 29-mm diameter, 65-mm length; isotropic spatial resolution 1–1.5-mm). Stenosis phantoms (n=3) consisted of polyvinyl chloride (PVC) tubes (8-mm inner diameter) prepared with centrally aligned cable binders to form a ~50% stenosis. A dedicated image reconstruction algorithm allowed precise tracking of endovascular instruments at 8 frames/s with a latency time of ~115 ms. A custom-made MPI-visible lacquer was used to manually label conventional guidewires, balloon catheters, and stainless steel balloon-expandable stents. Vascular stenoses were visualized by injecting a diluted SPIO tracer (ferucarbotran, 10 mmol iron/L) into the vessel phantoms. Balloon angioplasty and stent placement were performed by inflating balloon catheters and stent delivery balloons with diluted ferucarbotran.Results: After deployment of the stent, the markers on its ends were clearly visible. The applied lacquer markers were thin enough to not relevantly alter gliding properties of the devices while withstanding friction during the experiments. Placing an optimized flexible lacquer formulation on the preexisting radiopaque stent markers provided enough stability to withstand stent expansion. Final MPA confirmed successful stenosis treatment, facilitated by the disappearance of the lacquer markers on the stent due to differences in SPIO concentration. Thus, the in-stent lumen could be visualized without interference by the signal from the markers.Conclusion: Near real-time visualization of MPI-guided stenting of stenoses in a phantom model is feasible. Optimized MPI-visible markers can withstand the expansion process of stents. … (more)
- Is Part Of:
- Journal of endovascular therapy. Volume 26:Number 4(2019:Aug.)
- Journal:
- Journal of endovascular therapy
- Issue:
- Volume 26:Number 4(2019:Aug.)
- Issue Display:
- Volume 26, Issue 4 (2019)
- Year:
- 2019
- Volume:
- 26
- Issue:
- 4
- Issue Sort Value:
- 2019-0026-0004-0000
- Page Start:
- 512
- Page End:
- 519
- Publication Date:
- 2019-08
- Subjects:
- ferucarbotran -- in-stent lumen -- magnetic particle imaging -- percutaneous transluminal angioplasty -- phantom model -- real-time imaging -- stenosis -- stent -- superparamagnetic iron oxide nanoparticles
Blood-vessels -- Endoscopic surgery -- Periodicals
Angioscopy -- Periodicals
Intravenous catheterization -- Periodicals
Peripheral vascular diseases -- Treatment -- Periodicals
Vascular Surgical Procedures -- Periodicals
Angioscopy -- Periodicals
Catheterization, Peripheral -- Periodicals
Peripheral Vascular Diseases -- therapy -- Periodicals
Angioscopie
Maladies vasculaires périphériques
617.413 - Journal URLs:
- http://jet.sagepub.com/ ↗
http://www.jevt.org ↗
http://www.uk.sagepub.com ↗ - DOI:
- 10.1177/1526602819851202 ↗
- Languages:
- English
- ISSNs:
- 1526-6028
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 11450.xml