Advanced treatment planning using direct 4D optimisation for pencil-beam scanned particle therapy. (31st July 2017)
- Record Type:
- Journal Article
- Title:
- Advanced treatment planning using direct 4D optimisation for pencil-beam scanned particle therapy. (31st July 2017)
- Main Title:
- Advanced treatment planning using direct 4D optimisation for pencil-beam scanned particle therapy
- Authors:
- Bernatowicz, Kinga
Zhang, Ye
Perrin, Rosalind
Weber, Damien C
Lomax, Antony J - Abstract:
- Abstract: We report on development of a new four-dimensional (4D) optimisation approach for scanned proton beams, which incorporates both irregular motion patterns and the delivery dynamics of the treatment machine into the plan optimiser. Furthermore, we assess the effectiveness of this technique to reduce dose to critical structures in proximity to moving targets, while maintaining effective target dose homogeneity and coverage. The proposed approach has been tested using both a simulated phantom and a clinical liver cancer case, and allows for realistic 4D calculations and optimisation using irregular breathing patterns extracted from e.g. 4DCT-MRI (4D computed tomography-magnetic resonance imaging). 4D dose distributions resulting from our 4D optimisation can achieve almost the same quality as static plans, independent of the studied geometry/anatomy or selected motion (regular and irregular). Additionally, current implementation of the 4D optimisation approach requires less than 3 min to find the solution for a single field planned on 4DCT of a liver cancer patient. Although 4D optimisation allows for realistic calculations using irregular breathing patterns, it is very sensitive to variations from the planned motion. Based on a sensitivity analysis, target dose homogeneity comparable to static plans (D5-D95 <5%) has been found only for differences in amplitude of up to 1 mm, for changes in respiratory phase <200 ms and for changes in the breathing period of <20 msAbstract: We report on development of a new four-dimensional (4D) optimisation approach for scanned proton beams, which incorporates both irregular motion patterns and the delivery dynamics of the treatment machine into the plan optimiser. Furthermore, we assess the effectiveness of this technique to reduce dose to critical structures in proximity to moving targets, while maintaining effective target dose homogeneity and coverage. The proposed approach has been tested using both a simulated phantom and a clinical liver cancer case, and allows for realistic 4D calculations and optimisation using irregular breathing patterns extracted from e.g. 4DCT-MRI (4D computed tomography-magnetic resonance imaging). 4D dose distributions resulting from our 4D optimisation can achieve almost the same quality as static plans, independent of the studied geometry/anatomy or selected motion (regular and irregular). Additionally, current implementation of the 4D optimisation approach requires less than 3 min to find the solution for a single field planned on 4DCT of a liver cancer patient. Although 4D optimisation allows for realistic calculations using irregular breathing patterns, it is very sensitive to variations from the planned motion. Based on a sensitivity analysis, target dose homogeneity comparable to static plans (D5-D95 <5%) has been found only for differences in amplitude of up to 1 mm, for changes in respiratory phase <200 ms and for changes in the breathing period of <20 ms in comparison to the motions used during optimisation. As such, methods to robustly deliver 4D optimised plans employing 4D intensity-modulated delivery are discussed. … (more)
- Is Part Of:
- Physics in medicine & biology. Volume 62:Number 16(2017:Aug.)
- Journal:
- Physics in medicine & biology
- Issue:
- Volume 62:Number 16(2017:Aug.)
- Issue Display:
- Volume 62, Issue 16 (2017)
- Year:
- 2017
- Volume:
- 62
- Issue:
- 16
- Issue Sort Value:
- 2017-0062-0016-0000
- Page Start:
- 6595
- Page End:
- 6609
- Publication Date:
- 2017-07-31
- Subjects:
- 4D optimisation -- scanned proton therapy -- scanned particle therapy -- direct optimisation -- motion mitigation
Biophysics -- Periodicals
Medical physics -- Periodicals
610.153 - Journal URLs:
- http://ioppublishing.org/ ↗
http://iopscience.iop.org/0031-9155 ↗ - DOI:
- 10.1088/1361-6560/aa7ab8 ↗
- Languages:
- English
- ISSNs:
- 0031-9155
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 11427.xml