Adverse reaction to metal debris after Birmingham hip resurfacing arthroplasty. (4th May 2015)
- Record Type:
- Journal Article
- Title:
- Adverse reaction to metal debris after Birmingham hip resurfacing arthroplasty. (4th May 2015)
- Main Title:
- Adverse reaction to metal debris after Birmingham hip resurfacing arthroplasty
- Authors:
- Junnila, Mika
Seppänen, Matti
Mokka, Jari
Virolainen, Petri
Pölönen, Tuukka
Vahlberg, Tero
Mattila, Kimmo
Tuominen, Esa K J
Rantakokko, Juho
Äärimaa, Ville
Itälä, Ari
Mäkelä, Keijo T - Abstract:
- Abstract : Background and purpose — Concern has emerged about local soft-tissue reactions after hip resurfacing arthroplasty (HRA). The Birmingham Hip Resurfacing (BHR) was the most commonly used HRA device at our institution. We assessed the prevalence and risk factors for adverse reaction to metal debris (ARMD) with this device. Patients and methods — From 2003 to 2011, BHR was the most commonly used HRA device at our institution, with 249 implantations. We included 32 patients (24 of them men) who were operated with a BHR HRA during the period April 2004 to March 2007 (42 hips; 31 in men). The mean age of the patients was 59 (26–77) years. These patients underwent magnetic resonance imaging (MRI), serum metal ion measurements, the Oxford hip score questionnaire, and physical examination. The prevalence of ARMD was recorded, and risk factors for ARMD were assessed using logistic regression models. The mean follow-up time was 6.7 (2.4–8.8) years. Results — 6 patients had a definite ARMD (involving 9 of the 42 hips). 8 other patients (8 hips) had a probable ARMD. Thus, there was definite or probable ARMD in 17 of the 42 hips. 4 of 42 hips were revised for ARMD. Gender, bilateral metal-on-metal hip replacement and head size were not factors associated with ARMD. Interpretation — We found that HRA with the Birmingham Hip Resurfacing may be more dangerous than previously believed. We advise systematic follow-up of these patients using metal ion levels, MRI/ultrasound, andAbstract : Background and purpose — Concern has emerged about local soft-tissue reactions after hip resurfacing arthroplasty (HRA). The Birmingham Hip Resurfacing (BHR) was the most commonly used HRA device at our institution. We assessed the prevalence and risk factors for adverse reaction to metal debris (ARMD) with this device. Patients and methods — From 2003 to 2011, BHR was the most commonly used HRA device at our institution, with 249 implantations. We included 32 patients (24 of them men) who were operated with a BHR HRA during the period April 2004 to March 2007 (42 hips; 31 in men). The mean age of the patients was 59 (26–77) years. These patients underwent magnetic resonance imaging (MRI), serum metal ion measurements, the Oxford hip score questionnaire, and physical examination. The prevalence of ARMD was recorded, and risk factors for ARMD were assessed using logistic regression models. The mean follow-up time was 6.7 (2.4–8.8) years. Results — 6 patients had a definite ARMD (involving 9 of the 42 hips). 8 other patients (8 hips) had a probable ARMD. Thus, there was definite or probable ARMD in 17 of the 42 hips. 4 of 42 hips were revised for ARMD. Gender, bilateral metal-on-metal hip replacement and head size were not factors associated with ARMD. Interpretation — We found that HRA with the Birmingham Hip Resurfacing may be more dangerous than previously believed. We advise systematic follow-up of these patients using metal ion levels, MRI/ultrasound, and patient-reported outcome measures. … (more)
- Is Part Of:
- Acta orthopaedica. Volume 86:Number 3(2015)
- Journal:
- Acta orthopaedica
- Issue:
- Volume 86:Number 3(2015)
- Issue Display:
- Volume 86, Issue 3 (2015)
- Year:
- 2015
- Volume:
- 86
- Issue:
- 3
- Issue Sort Value:
- 2015-0086-0003-0000
- Page Start:
- 345
- Page End:
- 350
- Publication Date:
- 2015-05-04
- Subjects:
- Orthopedics -- Periodicals
616.7005 - Journal URLs:
- http://informahealthcare.com/loi/ort ↗
http://www.tandfonline.com/toc/iort20/current ↗
https://actaorthop.org/actao/index ↗
http://www.tandfonline.com/ ↗
http://journalsonline.tandf.co.uk/app/home/journal.asp?wasp=65168817ff044fea9c5b577f1cfe2186&referrer=parent&backto=linkingpublicationresults, 1:113260, 1 ↗ - DOI:
- 10.3109/17453674.2014.1004015 ↗
- Languages:
- English
- ISSNs:
- 1745-3674
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 0642.055000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 11407.xml