Localised hyperthermia in rodent models using an MRI-compatible high-intensity focused ultrasound system. (17th November 2015)
- Record Type:
- Journal Article
- Title:
- Localised hyperthermia in rodent models using an MRI-compatible high-intensity focused ultrasound system. (17th November 2015)
- Main Title:
- Localised hyperthermia in rodent models using an MRI-compatible high-intensity focused ultrasound system
- Authors:
- Bing, Chenchen
Nofiele, Joris
Staruch, Robert
Ladouceur-Wodzak, Michelle
Chatzinoff, Yonatan
Ranjan, Ashish
Chopra, Rajiv - Abstract:
- Abstract: Purpose : Localised hyperthermia in rodent studies is challenging due to the small target size. This study describes the development and characterisation of an MRI-compatible high-intensity focused ultrasound (HIFU) system to perform localised mild hyperthermia treatments in rodent models. Material and methods : The hyperthermia platform consisted of an MRI-compatible small animal HIFU system, focused transducers with sector-vortex lenses, a custom-made receive coil, and means to maintain systemic temperatures of rodents. The system was integrated into a 3T MR imager. Control software was developed to acquire images, process temperature maps, and adjust output power using a proportional-integral-derivative feedback control algorithm. Hyperthermia exposures were performed in tissue-mimicking phantoms and in a rodent model ( n = 9). During heating, an ROI was assigned in the heated region for temperature control and the target temperature was 42 °C; 30 min mild hyperthermia treatment followed by a 10-min cooling procedure was performed on each animal. Results : 3D-printed sector-vortex lenses were successful at creating annular focal regions which enables customisation of the heating volume. Localised mild hyperthermia performed in rats produced a mean ROI temperature of 42.1 ± 0.3 °C. The T10 and T90 percentiles were 43.2 ± 0.4 °C and 41.0 ± 0.3 °C, respectively. For a 30-min treatment, the mean time duration between 41–45 °C was 31.1 min within the ROI.Abstract: Purpose : Localised hyperthermia in rodent studies is challenging due to the small target size. This study describes the development and characterisation of an MRI-compatible high-intensity focused ultrasound (HIFU) system to perform localised mild hyperthermia treatments in rodent models. Material and methods : The hyperthermia platform consisted of an MRI-compatible small animal HIFU system, focused transducers with sector-vortex lenses, a custom-made receive coil, and means to maintain systemic temperatures of rodents. The system was integrated into a 3T MR imager. Control software was developed to acquire images, process temperature maps, and adjust output power using a proportional-integral-derivative feedback control algorithm. Hyperthermia exposures were performed in tissue-mimicking phantoms and in a rodent model ( n = 9). During heating, an ROI was assigned in the heated region for temperature control and the target temperature was 42 °C; 30 min mild hyperthermia treatment followed by a 10-min cooling procedure was performed on each animal. Results : 3D-printed sector-vortex lenses were successful at creating annular focal regions which enables customisation of the heating volume. Localised mild hyperthermia performed in rats produced a mean ROI temperature of 42.1 ± 0.3 °C. The T10 and T90 percentiles were 43.2 ± 0.4 °C and 41.0 ± 0.3 °C, respectively. For a 30-min treatment, the mean time duration between 41–45 °C was 31.1 min within the ROI. Conclusions : The MRI-compatible HIFU system was successfully adapted to perform localised mild hyperthermia treatment in rodent models. A target temperature of 42 °C was well-maintained in a rat thigh model for 30 min. … (more)
- Is Part Of:
- International journal of hyperthermia. Volume 31:Number 8(2015)
- Journal:
- International journal of hyperthermia
- Issue:
- Volume 31:Number 8(2015)
- Issue Display:
- Volume 31, Issue 8 (2015)
- Year:
- 2015
- Volume:
- 31
- Issue:
- 8
- Issue Sort Value:
- 2015-0031-0008-0000
- Page Start:
- 813
- Page End:
- 822
- Publication Date:
- 2015-11-17
- Subjects:
- Heat-targeted drug delivery -- mild hyperthermia -- MR-HIFU -- nanoparticles -- non-invasive thermometry
Thermotherapy -- Periodicals
615.832 - Journal URLs:
- http://informahealthcare.com/loi/hth ↗
http://www.tandf.co.uk/journals/titles/02656736.asp ↗
http://informahealthcare.com ↗ - DOI:
- 10.3109/02656736.2015.1094833 ↗
- Languages:
- English
- ISSNs:
- 0265-6736
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 4542.297000
British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 11396.xml