High dose‐per‐pulse electron beam dosimetry: Usability and dose‐rate independence of EBT3 Gafchromic films. Issue 2 (8th February 2017)
- Record Type:
- Journal Article
- Title:
- High dose‐per‐pulse electron beam dosimetry: Usability and dose‐rate independence of EBT3 Gafchromic films. Issue 2 (8th February 2017)
- Main Title:
- High dose‐per‐pulse electron beam dosimetry: Usability and dose‐rate independence of EBT3 Gafchromic films
- Authors:
- Jaccard, Maud
Petersson, Kristoffer
Buchillier, Thierry
Germond, Jean‐François
Durán, Maria Teresa
Vozenin, Marie‐Catherine
Bourhis, Jean
Bochud, François O.
Bailat, Claude - Abstract:
- Abstract : Purpose: The aim of this study was to assess the suitability of Gafchromic EBT3 films for reference dose measurements in the beam of a prototype high dose‐per‐pulse linear accelerator (linac), capable of delivering electron beams with a mean dose‐rate ( Ḋ m ) ranging from 0.07 to 3000 Gy/s and a dose‐rate in pulse ( Ḋ p ) of up to 8 × 10 6 Gy/s. To do this, we evaluated the overall uncertainties in EBT3 film dosimetry as well as the energy and dose‐rate dependence of their response. Material and methods: Our dosimetric system was composed of EBT3 Gafchromic films in combination with a flatbed scanner and was calibrated against an ionization chamber traceable to primary standard. All sources of uncertainties in EBT3 dosimetry were carefully analyzed using irradiations at a clinical radiotherapy linac. Energy dependence was investigated with the same machine by acquiring and comparing calibration curves for three different beam energies (4, 8 and 12 MeV), for doses between 0.25 and 30 Gy. Ḋ m dependence was studied at the clinical linac by changing the pulse repetition frequency ( f ) of the beam in order to vary Ḋ m between 0.55 and 4.40 Gy/min, while Ḋ p dependence was probed at the prototype machine for Ḋ p ranging from 7 × 10 3 to 8 × 10 6 Gy/s. Ḋ p dependence was first determined by studying the correlation between the dose measured by films and the charge of electrons measured at the exit of the machine by an induction torus. Furthermore, we compared dosesAbstract : Purpose: The aim of this study was to assess the suitability of Gafchromic EBT3 films for reference dose measurements in the beam of a prototype high dose‐per‐pulse linear accelerator (linac), capable of delivering electron beams with a mean dose‐rate ( Ḋ m ) ranging from 0.07 to 3000 Gy/s and a dose‐rate in pulse ( Ḋ p ) of up to 8 × 10 6 Gy/s. To do this, we evaluated the overall uncertainties in EBT3 film dosimetry as well as the energy and dose‐rate dependence of their response. Material and methods: Our dosimetric system was composed of EBT3 Gafchromic films in combination with a flatbed scanner and was calibrated against an ionization chamber traceable to primary standard. All sources of uncertainties in EBT3 dosimetry were carefully analyzed using irradiations at a clinical radiotherapy linac. Energy dependence was investigated with the same machine by acquiring and comparing calibration curves for three different beam energies (4, 8 and 12 MeV), for doses between 0.25 and 30 Gy. Ḋ m dependence was studied at the clinical linac by changing the pulse repetition frequency ( f ) of the beam in order to vary Ḋ m between 0.55 and 4.40 Gy/min, while Ḋ p dependence was probed at the prototype machine for Ḋ p ranging from 7 × 10 3 to 8 × 10 6 Gy/s. Ḋ p dependence was first determined by studying the correlation between the dose measured by films and the charge of electrons measured at the exit of the machine by an induction torus. Furthermore, we compared doses from the films to independently calibrated thermo‐luminescent dosimeters (TLD) that have been reported as being dose‐rate independent up to such high dose‐rates. Results: We report that uncertainty below 4% ( k = 2) can be achieved in the dose range between 3 and 17 Gy. Results also demonstrated that EBT3 films did not display any detectable energy dependence for electron beam energies between 4 and 12 MeV. No Ḋ m dependence was found either. In addition, we obtained excellent consistency between films and TLDs over the entire Ḋ p range attainable at the prototype linac confirming the absence of any dose‐rate dependence within the investigated range (7 × 10 3 to 8 × 10 6 Gy/s). This aspect was further corroborated by the linear relationship between the dose‐per‐pulse ( D p ) measured by films and the charge per pulse ( C p ) measured at the prototype linac exit. Conclusion: Our study shows that the use of EBT3 Gafchromic films can be extended to reference dosimetry in pulsed electron beams with a very high dose rate. The measurement results are associated with an overall uncertainty below 4% ( k = 2) and are dose‐rate and energy independent. … (more)
- Is Part Of:
- Medical physics. Volume 44:Issue 2(2017)
- Journal:
- Medical physics
- Issue:
- Volume 44:Issue 2(2017)
- Issue Display:
- Volume 44, Issue 2 (2017)
- Year:
- 2017
- Volume:
- 44
- Issue:
- 2
- Issue Sort Value:
- 2017-0044-0002-0000
- Page Start:
- 725
- Page End:
- 735
- Publication Date:
- 2017-02-08
- Subjects:
- dose‐rate dependence -- EBT3 Gafchromic films -- energy dependence -- high dose‐per‐pulse -- high dose‐rate
Medical physics -- Periodicals
Medical physics
Geneeskunde
Natuurkunde
Toepassingen
Biophysics
Periodicals
Periodicals
Electronic journals
610.153 - Journal URLs:
- http://scitation.aip.org/content/aapm/journal/medphys ↗
https://aapm.onlinelibrary.wiley.com/journal/24734209 ↗
http://www.aip.org/ ↗ - DOI:
- 10.1002/mp.12066 ↗
- Languages:
- English
- ISSNs:
- 0094-2405
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 5531.130000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 11308.xml