Evaluation of predicted knee function for component malrotation in total knee arthroplasty. (February 2017)
- Record Type:
- Journal Article
- Title:
- Evaluation of predicted knee function for component malrotation in total knee arthroplasty. (February 2017)
- Main Title:
- Evaluation of predicted knee function for component malrotation in total knee arthroplasty
- Authors:
- Vanheule, Valentine
Delport, Hendrik Pieter
Andersen, Michael Skipper
Scheys, Lennart
Wirix-Speetjens, Roel
Jonkers, Ilse
Victor, Jan
Vander Sloten, Jos - Abstract:
- Highlights: Multi-body model quantitatively predicted subject-specific knee behaviour after TKA. Experimental testing of impact of TKA malrotation on ligament lengths and motion. Presented model could distinguish effect of TKA malrotation on knee behaviour. Abstract: Soft-tissue balancing for total knee arthroplasty (TKA) remains subjective and highly dependent on surgical expertise. Pre-operative planning may support the clinician in taking decisions by integrating subject-specific computer models that predict functional outcome. However, validation of these models is essential before they can be applied in clinical practice. The aim of this study was to evaluate a knee modelling workflow by comparing experimental cadaveric measures to model-based kinematics and ligament length changes. Subject-specific models for three cadaveric knees were constructed from medical images. The implanted knees were mounted onto a mechanical rig to perform squatting, measuring kinematics and ligament length changes with optical markers and extensometers. Coronal malrotation was introduced using tibial inserts with a built-in slope. The model output agreed well with the experiment in all alignment conditions. Kinematic behaviour showed an average RMSE of less than 2.7 mm and 2.3° for translations and rotations. The average RMSE was below 2.5% for all ligaments. These results show that the presented model can quantitatively predict subject-specific knee behaviour following TKA, allowingHighlights: Multi-body model quantitatively predicted subject-specific knee behaviour after TKA. Experimental testing of impact of TKA malrotation on ligament lengths and motion. Presented model could distinguish effect of TKA malrotation on knee behaviour. Abstract: Soft-tissue balancing for total knee arthroplasty (TKA) remains subjective and highly dependent on surgical expertise. Pre-operative planning may support the clinician in taking decisions by integrating subject-specific computer models that predict functional outcome. However, validation of these models is essential before they can be applied in clinical practice. The aim of this study was to evaluate a knee modelling workflow by comparing experimental cadaveric measures to model-based kinematics and ligament length changes. Subject-specific models for three cadaveric knees were constructed from medical images. The implanted knees were mounted onto a mechanical rig to perform squatting, measuring kinematics and ligament length changes with optical markers and extensometers. Coronal malrotation was introduced using tibial inserts with a built-in slope. The model output agreed well with the experiment in all alignment conditions. Kinematic behaviour showed an average RMSE of less than 2.7 mm and 2.3° for translations and rotations. The average RMSE was below 2.5% for all ligaments. These results show that the presented model can quantitatively predict subject-specific knee behaviour following TKA, allowing evaluation of implant alignment in terms of kinematics and ligament length changes. In future work, the model will be used to evaluate subject-specific implant position based on ligament behaviour. … (more)
- Is Part Of:
- Medical engineering & physics. Volume 40(2017)
- Journal:
- Medical engineering & physics
- Issue:
- Volume 40(2017)
- Issue Display:
- Volume 40, Issue 2017 (2017)
- Year:
- 2017
- Volume:
- 40
- Issue:
- 2017
- Issue Sort Value:
- 2017-0040-2017-0000
- Page Start:
- 56
- Page End:
- 64
- Publication Date:
- 2017-02
- Subjects:
- Subject-specific -- Alignment -- Soft tissue balancing -- Kinematic knee rig -- In vitro
Biomedical engineering -- Periodicals
Biomedical Engineering -- Periodicals
Physics -- Periodicals
Génie biomédical -- Périodiques
Biomedical engineering
Electronic journals
Periodicals
610.28 - Journal URLs:
- http://www.medengphys.com ↗
http://www.sciencedirect.com/science/journal/13504533 ↗
http://www.clinicalkey.com/dura/browse/journalIssue/13504533 ↗
http://www.clinicalkey.com.au/dura/browse/journalIssue/13504533 ↗
http://www.elsevier.com/journals ↗ - DOI:
- 10.1016/j.medengphy.2016.12.001 ↗
- Languages:
- English
- ISSNs:
- 1350-4533
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 5527.323000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 11220.xml