Goldmann Applanation Tonometer Versus Ocular Response Analyzer for Measuring Intraocular Pressure After Descemet Stripping Automated Endothelial Keratoplasty. Issue 11 (November 2018)
- Record Type:
- Journal Article
- Title:
- Goldmann Applanation Tonometer Versus Ocular Response Analyzer for Measuring Intraocular Pressure After Descemet Stripping Automated Endothelial Keratoplasty. Issue 11 (November 2018)
- Main Title:
- Goldmann Applanation Tonometer Versus Ocular Response Analyzer for Measuring Intraocular Pressure After Descemet Stripping Automated Endothelial Keratoplasty
- Authors:
- Feizi, Sepehr
Faramarzi, Amir
Masoudi, Ali
Azari, Amir A.
Veisi, Amirreza - Abstract:
- Abstract : Purpose: To evaluate agreement between IOP measured using the ocular response analyzer (ORA) versus using the Goldmann applanation tonometer (GAT) in patients treated by Descemet stripping automated endothelial keratoplasty (DSAEK). Methods: This prospective, case–control study comprised 38 normal eyes and 37 post-DSAEK eyes. IOP was measured using the GAT (IOPGAT ). The ORA was used to measure corneal biomechanics and cornea-compensated IOP (IOPCC ). Corneal hysteresis (CH), corneal resistance factor, and IOP measurements of the 2 groups were compared. Intertonometer agreement was evaluated in each study group using Bland–Altman plots and 95% limits of agreement. Univariate analysis was used to evaluate the effect of independent factors on the difference between the 2 IOP readings (ΔIOP). Results: The IOPCC and IOPGAT readings showed no significant intergroup differences. IOPCC was significantly higher than IOPGAT, with a mean difference of 3.0 ± 3.3 mm Hg in the normal group ( P < 0.001) and 4.5 ± 3.1 mm Hg in the DSAEK group ( P < 0.001). The 95% limits of agreement for the IOPCC and IOPGAT was between −3.4 and 9.5 mm Hg and between −1.6 and 10.6 mm Hg in the normal and DSAEK groups, respectively. CH (r = −0.57, P < 0.001) and donor thickness (r = 0.81, P = 0.04) were identified as the main predictors of ΔIOP after DSAEK. ΔIOP did not vary significantly with the central corneal thickness or corneal resistance factor. Conclusions: The IOPGAT and IOPCC wereAbstract : Purpose: To evaluate agreement between IOP measured using the ocular response analyzer (ORA) versus using the Goldmann applanation tonometer (GAT) in patients treated by Descemet stripping automated endothelial keratoplasty (DSAEK). Methods: This prospective, case–control study comprised 38 normal eyes and 37 post-DSAEK eyes. IOP was measured using the GAT (IOPGAT ). The ORA was used to measure corneal biomechanics and cornea-compensated IOP (IOPCC ). Corneal hysteresis (CH), corneal resistance factor, and IOP measurements of the 2 groups were compared. Intertonometer agreement was evaluated in each study group using Bland–Altman plots and 95% limits of agreement. Univariate analysis was used to evaluate the effect of independent factors on the difference between the 2 IOP readings (ΔIOP). Results: The IOPCC and IOPGAT readings showed no significant intergroup differences. IOPCC was significantly higher than IOPGAT, with a mean difference of 3.0 ± 3.3 mm Hg in the normal group ( P < 0.001) and 4.5 ± 3.1 mm Hg in the DSAEK group ( P < 0.001). The 95% limits of agreement for the IOPCC and IOPGAT was between −3.4 and 9.5 mm Hg and between −1.6 and 10.6 mm Hg in the normal and DSAEK groups, respectively. CH (r = −0.57, P < 0.001) and donor thickness (r = 0.81, P = 0.04) were identified as the main predictors of ΔIOP after DSAEK. ΔIOP did not vary significantly with the central corneal thickness or corneal resistance factor. Conclusions: The IOPGAT and IOPCC were significantly different, which indicates that the GAT and ORA cannot be used interchangeably for measuring IOP after DSAEK. The difference between these 2 measurements was primarily affected by the variations in CH and donor thickness. … (more)
- Is Part Of:
- Cornea. Volume 37:Issue 11(2018)
- Journal:
- Cornea
- Issue:
- Volume 37:Issue 11(2018)
- Issue Display:
- Volume 37, Issue 11 (2018)
- Year:
- 2018
- Volume:
- 37
- Issue:
- 11
- Issue Sort Value:
- 2018-0037-0011-0000
- Page Start:
- Page End:
- Publication Date:
- 2018-11
- Subjects:
- Descemet stripping automated endothelial keratoplasty -- intraocular pressure -- Goldmann applanation tonometer -- ocular response analyzer -- cornea-compensated intraocular pressure
Cornea -- Periodicals
Cornea -- Periodicals
Cornée -- Périodiques
617.719 - Journal URLs:
- http://journals.lww.com/corneajrnl/pages/default.aspx ↗
http://journals.lww.com ↗ - DOI:
- 10.1097/ICO.0000000000001660 ↗
- Languages:
- English
- ISSNs:
- 0277-3740
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 3470.927500
British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 11226.xml