International Analysis of LVAD Point-of-Care Versus Plasma INR: A Multicenter Study. Issue 6 (November 2018)
- Record Type:
- Journal Article
- Title:
- International Analysis of LVAD Point-of-Care Versus Plasma INR: A Multicenter Study. Issue 6 (November 2018)
- Main Title:
- International Analysis of LVAD Point-of-Care Versus Plasma INR
- Authors:
- Schettle, Sarah
Schlöglhofer, Thomas
Zimpfer, Daniel
Schima, Heinrich
Kaufmann, Friedrich
Salimbangon, Anthony
Wolforth, Jennifer
Schroeder, Sarah
Hards, Rachel
Ritchie, Brianne
Staley, Linda
Pereira, Naveen
Kushwaha, Sudhir
Maltais, Simon
Stulak, John - Abstract:
- Abstract : International normalized ratio (INR) measurements manage risks of bleeding and clotting sequelae in the population with left ventricular assist device (LVAD). Studies suggest that point-of-care (POC) INR does not significantly differ from plasma INR (P-INR) values in other patient populations, although this has not been validated in a multicenter LVAD study. We sought to determine whether POC-INR and P-INR values differ significantly in the LVAD population and reviewed patients with permanent LVAD implantation at seven institutions internationally. Our cohort comprised of 279 paired POC-INR and P-INR checks in patients supported on average 630 ± 598 days postimplant with predominately HeartWare, HeartMate II, and HeartMate III devices. The population averaged 57.9 years of age, and there were 86.7% male. We found no statistically significant difference between POC-INR and P-INR values. International normalized ratio accuracy correlated weakly with the time between INR measurements ( p < 0.001). When the time difference was less than 4 hours, the difference between INR pairs was significantly lower than measurements greater than 8 hours ( p = 0.006). There was no statistically significant difference when comparing paired INR values and time after implant to INR check ( p = 0.43), age ( p = 0.12), known coagulopathy ( p = 0.12), bleeding history ( p = 0.22), or thrombosis history ( p = 0.34). This is the first large multicenter international study comparing POC-INRAbstract : International normalized ratio (INR) measurements manage risks of bleeding and clotting sequelae in the population with left ventricular assist device (LVAD). Studies suggest that point-of-care (POC) INR does not significantly differ from plasma INR (P-INR) values in other patient populations, although this has not been validated in a multicenter LVAD study. We sought to determine whether POC-INR and P-INR values differ significantly in the LVAD population and reviewed patients with permanent LVAD implantation at seven institutions internationally. Our cohort comprised of 279 paired POC-INR and P-INR checks in patients supported on average 630 ± 598 days postimplant with predominately HeartWare, HeartMate II, and HeartMate III devices. The population averaged 57.9 years of age, and there were 86.7% male. We found no statistically significant difference between POC-INR and P-INR values. International normalized ratio accuracy correlated weakly with the time between INR measurements ( p < 0.001). When the time difference was less than 4 hours, the difference between INR pairs was significantly lower than measurements greater than 8 hours ( p = 0.006). There was no statistically significant difference when comparing paired INR values and time after implant to INR check ( p = 0.43), age ( p = 0.12), known coagulopathy ( p = 0.12), bleeding history ( p = 0.22), or thrombosis history ( p = 0.34). This is the first large multicenter international study comparing POC-INR and P-INR measurements in patients with LVAD and found no statistically significant difference between either methods, particularly when measured within less than 4 hours of each other. … (more)
- Is Part Of:
- ASAIO journal. Volume 64:Issue 6(2018)
- Journal:
- ASAIO journal
- Issue:
- Volume 64:Issue 6(2018)
- Issue Display:
- Volume 64, Issue 6 (2018)
- Year:
- 2018
- Volume:
- 64
- Issue:
- 6
- Issue Sort Value:
- 2018-0064-0006-0000
- Page Start:
- Page End:
- Publication Date:
- 2018-11
- Subjects:
- LVAD -- home INR -- LVAD warfarin -- LVAD anticoagulation -- LVAD point of care INR
Artificial organs -- Periodicals
617 - Journal URLs:
- http://journals.lww.com/asaiojournal/pages/default.aspx ↗
http://journals.lww.com ↗ - DOI:
- 10.1097/MAT.0000000000000845 ↗
- Languages:
- English
- ISSNs:
- 1058-2916
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 1738.840500
British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 11216.xml