Experimental aspects of buoyancy correction in measuring reliable high-pressure excess adsorption isotherms using the gravimetric method. (21st November 2017)
- Record Type:
- Journal Article
- Title:
- Experimental aspects of buoyancy correction in measuring reliable high-pressure excess adsorption isotherms using the gravimetric method. (21st November 2017)
- Main Title:
- Experimental aspects of buoyancy correction in measuring reliable high-pressure excess adsorption isotherms using the gravimetric method
- Authors:
- Nguyen, Huong Giang T
Horn, Jarod C
Thommes, Matthias
van Zee, Roger D
Espinal, Laura - Abstract:
- Abstract: Addressing reproducibility issues in adsorption measurements is critical to accelerating the path to discovery of new industrial adsorbents and to understanding adsorption processes. A National Institute of Standards and Technology Reference Material, RM 8852 (ammonium ZSM-5 zeolite), and two gravimetric instruments with asymmetric two-beam balances were used to measure high-pressure adsorption isotherms. This work demonstrates how common approaches to buoyancy correction, a key factor in obtaining the mass change due to surface excess gas uptake from the apparent mass change, can impact the adsorption isotherm data. Three different approaches to buoyancy correction were investigated and applied to the subcritical CO2 and supercritical N2 adsorption isotherms at 293 K. It was observed that measuring a collective volume for all balance components for the buoyancy correction (helium method) introduces an inherent bias in temperature partition when there is a temperature gradient (i.e. analysis temperature is not equal to instrument air bath temperature). We demonstrate that a blank subtraction is effective in mitigating the biases associated with temperature partitioning, instrument calibration, and the determined volumes of the balance components. In general, the manual and subtraction methods allow for better treatment of the temperature gradient during buoyancy correction. From the study, best practices specific to asymmetric two-beam balances and more generalAbstract: Addressing reproducibility issues in adsorption measurements is critical to accelerating the path to discovery of new industrial adsorbents and to understanding adsorption processes. A National Institute of Standards and Technology Reference Material, RM 8852 (ammonium ZSM-5 zeolite), and two gravimetric instruments with asymmetric two-beam balances were used to measure high-pressure adsorption isotherms. This work demonstrates how common approaches to buoyancy correction, a key factor in obtaining the mass change due to surface excess gas uptake from the apparent mass change, can impact the adsorption isotherm data. Three different approaches to buoyancy correction were investigated and applied to the subcritical CO2 and supercritical N2 adsorption isotherms at 293 K. It was observed that measuring a collective volume for all balance components for the buoyancy correction (helium method) introduces an inherent bias in temperature partition when there is a temperature gradient (i.e. analysis temperature is not equal to instrument air bath temperature). We demonstrate that a blank subtraction is effective in mitigating the biases associated with temperature partitioning, instrument calibration, and the determined volumes of the balance components. In general, the manual and subtraction methods allow for better treatment of the temperature gradient during buoyancy correction. From the study, best practices specific to asymmetric two-beam balances and more general recommendations for measuring isotherms far from critical temperatures using gravimetric instruments are offered. … (more)
- Is Part Of:
- Measurement science & technology. Volume 28:Number 12(2017:Dec.)
- Journal:
- Measurement science & technology
- Issue:
- Volume 28:Number 12(2017:Dec.)
- Issue Display:
- Volume 28, Issue 12 (2017)
- Year:
- 2017
- Volume:
- 28
- Issue:
- 12
- Issue Sort Value:
- 2017-0028-0012-0000
- Page Start:
- Page End:
- Publication Date:
- 2017-11-21
- Subjects:
- high-pressure adsorption isotherm -- excess adsorption isotherm -- gravimetric method -- buoyancy correction -- blank correction -- RM 8852 -- ZSM-5
Physical measurements -- Periodicals
Scientific apparatus and instruments -- Periodicals
Equipment and Supplies -- Periodicals
Science -- instrumentation -- Periodicals
Technology -- instrumentation -- Periodicals
Mesures physiques -- Périodiques
Physical measurements
Scientific apparatus and instruments
Periodicals
502.87 - Journal URLs:
- http://iopscience.iop.org/0957-0233/ ↗
http://www.iop.org/Journals/mt ↗
http://ioppublishing.org/ ↗ - DOI:
- 10.1088/1361-6501/aa8f83 ↗
- Languages:
- English
- ISSNs:
- 0957-0233
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File: