- Characterisation of porous knitted titanium for replacement of intervertebral disc nucleus pulposus
- Authors:
- Tendulkar, Gauri
Sreekumar, Vrinda
Rupp, Frank
Teotia, Arun
Athanasopulu, Kiriaki
Kemkemer, Ralf
Buck, Alfred
Buck, Alfred
Kaps, Hans-Peter
Geis-Gerstorfer, Jürgen
Kumar, Ashok
Nussler, Andreas - Abstract:
- Abstract Effective restoration of human intervertebral disc degeneration is challenged by numerous limitations of the currently available spinal fusion and arthroplasty treatment strategies. Consequently, use of artificial biomaterial implant is gaining attention as a potential therapeutic strategy. Our study is aimed at investigating and characterizing a novel knitted titanium (Ti6Al4V) implant for the replacement of nucleus pulposus to treat early stages of chronic intervertebral disc degeneration. Specific knitted geometry of the scaffold with a porosity of 67.67 ± 0.824% was used to overcome tissue integration failures. Furthermore, to improve the wear resistance without impairing original mechanical strength, electro-polishing step was employed. Electro-polishing treatment changed a surface roughness from 15.22 ± 3.28 to 4.35 ± 0.87 µm without affecting its wettability which remained at 81.03 ± 8.5°. Subsequently, cellular responses of human mesenchymal stem cells (SCP1 cell line) and human primary chondrocytes were investigated which showed positive responses in terms of adherence and viability. Surface wettability was further enhanced to super hydrophilic nature by oxygen plasma treatment, which eventually caused substantial increase in the proliferation of SCP1 cells and primary chondrocytes. Our study implies that owing to scaffolds physicochemical and biocompatible properties, it could improve the clinical performance of nucleus pulposus replacement.
- Is Part Of:
- Scientific reports. Volume 7:Issue 1(2017)
- Journal:
- Scientific reports
- Issue:
- Volume 7:Issue 1(2017)
- Issue Display:
- Volume 7, Issue 1 (2017)
- Year:
- 2017
- Volume:
- 7
- Issue:
- 1
- Issue Sort Value:
- 2017-0007-0001-0000
- Page Start:
- 1
- Page End:
- 12
- Publication Date:
- 2017-12
- Subjects:
- Natural history -- Research -- Periodicals
Biology -- Research -- Periodicals
Physical sciences -- Research -- Periodicals
Earth sciences -- Research -- Periodicals
Environmental sciences -- Research -- Periodicals
502.85 - Journal URLs:
- http://www.nature.com/ ↗
http://www.nature.com/srep/index.html ↗ - DOI:
- 10.1038/s41598-017-16863-8 ↗
- Languages:
- English
- ISSNs:
- 2045-2322
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 11176.xml