Mechanical fatigue behaviour of different lengths screw‐retained restorations connected to two designs prosthetic connection level. (12th May 2019)
- Record Type:
- Journal Article
- Title:
- Mechanical fatigue behaviour of different lengths screw‐retained restorations connected to two designs prosthetic connection level. (12th May 2019)
- Main Title:
- Mechanical fatigue behaviour of different lengths screw‐retained restorations connected to two designs prosthetic connection level
- Authors:
- Pérez León, Patricia
Bartolomé, José F.
Lombardía, Carlos
Pradíes, Guillermo - Abstract:
- Abstract: The aim of this work is to compare the mechanical fatigue behaviour of different crown heights of single crown implant restorations made over two different internal hexagonal titanium dental implant connection levels. Two different groups of internal hexagonal connection were studied: (a) TL group (N = 26) included implants with a transmucosal connection (12.2 mm height and 4.8 mm width) and (b) BL group (N = 26) composed by implants at a bone level connection (10 mm height and 3.8 mm width). Two subgroups (N = 13) were established for each connection level according to crown heights (TL10 = 10 mm, TL15 = 15 mm, BL12 = 12 mm, BL17 = 17 mm). Dynamic load tests were carried out according to ISO Norm 14801. Failed samples and fracture surfaces were analysed with an optical and scanning electron microscope (SEM). The fatigue limit and the fatigue strength degradation (fatigue strength exponent) decreases with increasing crown height (5 mm) in both groups of internal hexagonal titanium dental implant connection levels. Consequently, the mechanical integrity of internal hexagonal dental implants varies with the height of the crown. An increase of 5 millimetres in the crown height appears to significantly decrease the fatigue life of both types of dental implants.
- Is Part Of:
- Journal of oral rehabilitation. Volume 46:Number 8(2019:Aug.)
- Journal:
- Journal of oral rehabilitation
- Issue:
- Volume 46:Number 8(2019:Aug.)
- Issue Display:
- Volume 46, Issue 8 (2019)
- Year:
- 2019
- Volume:
- 46
- Issue:
- 8
- Issue Sort Value:
- 2019-0046-0008-0000
- Page Start:
- 747
- Page End:
- 755
- Publication Date:
- 2019-05-12
- Subjects:
- crown height -- failure analysis -- fatigue -- implant level -- single implant
Dentistry -- Periodicals
Prosthodontics -- Periodicals
617 - Journal URLs:
- http://www.blackwell-synergy.com/member/institutions/issuelist.asp?journal=jor ↗
http://onlinelibrary.wiley.com/ ↗ - DOI:
- 10.1111/joor.12809 ↗
- Languages:
- English
- ISSNs:
- 0305-182X
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 5026.440000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 11118.xml