Intraoperative Error Propagation in 3-Dimensional Spinal Navigation From Nonsegmental Registration: A Prospective Cadaveric and Clinical Study. Issue 5 (August 2019)
- Record Type:
- Journal Article
- Title:
- Intraoperative Error Propagation in 3-Dimensional Spinal Navigation From Nonsegmental Registration: A Prospective Cadaveric and Clinical Study. Issue 5 (August 2019)
- Main Title:
- Intraoperative Error Propagation in 3-Dimensional Spinal Navigation From Nonsegmental Registration: A Prospective Cadaveric and Clinical Study
- Authors:
- Guha, Daipayan
Jakubovic, Raphael
Gupta, Shaurya
Fehlings, Michael G.
Mainprize, Todd G.
Yee, Albert
Yang, Victor X. D. - Abstract:
- Study Design: Prospective pre-clinical and clinical cohort study. Objectives: Current spinal navigation systems rely on a dynamic reference frame (DRF) for image-to-patient registration and tool tracking. Working distant to a DRF may generate inaccuracy. Here we quantitate predictors of navigation error as a function of distance from the registered vertebral level, and from intersegmental mobility due to surgical manipulation and patient respiration. Methods: Navigation errors from working distant to the registered level, and from surgical manipulation, were quantified in 4 human cadavers. The 3-dimensional (3D) position of a tracked tool tip at 0 to 5 levels from the DRF, and during targeting of pedicle screw tracts, was captured in real-time by an optical navigation system. Respiration-induced vertebral motion was quantified from 10 clinical cases of open posterior instrumentation. The 3D position of a custom spinous-process clamp was tracked over 12 respiratory cycles. Results: An increase in mean 3D navigation error of ≥2 mm was observed at ≥2 levels from the DRF in the cervical and lumbar spine. Mean ± SD displacement due to surgical manipulation was 1.55 ± 1.13 mm in 3D across all levels, ≥2 mm in 17.4%, 19.2%, and 38.5% of levels in the cervical, thoracic, and lumbar spine, respectively. Mean ± SD respiration-induced 3D motion was 1.96 ± 1.32 mm, greatest in the lower thoracic spine ( P < .001). Tidal volume and positive end-expiratory pressure correlated positivelyStudy Design: Prospective pre-clinical and clinical cohort study. Objectives: Current spinal navigation systems rely on a dynamic reference frame (DRF) for image-to-patient registration and tool tracking. Working distant to a DRF may generate inaccuracy. Here we quantitate predictors of navigation error as a function of distance from the registered vertebral level, and from intersegmental mobility due to surgical manipulation and patient respiration. Methods: Navigation errors from working distant to the registered level, and from surgical manipulation, were quantified in 4 human cadavers. The 3-dimensional (3D) position of a tracked tool tip at 0 to 5 levels from the DRF, and during targeting of pedicle screw tracts, was captured in real-time by an optical navigation system. Respiration-induced vertebral motion was quantified from 10 clinical cases of open posterior instrumentation. The 3D position of a custom spinous-process clamp was tracked over 12 respiratory cycles. Results: An increase in mean 3D navigation error of ≥2 mm was observed at ≥2 levels from the DRF in the cervical and lumbar spine. Mean ± SD displacement due to surgical manipulation was 1.55 ± 1.13 mm in 3D across all levels, ≥2 mm in 17.4%, 19.2%, and 38.5% of levels in the cervical, thoracic, and lumbar spine, respectively. Mean ± SD respiration-induced 3D motion was 1.96 ± 1.32 mm, greatest in the lower thoracic spine ( P < .001). Tidal volume and positive end-expiratory pressure correlated positively with increased vertebral displacement. Conclusions: Vertebral motion is unaccounted for during image-guided surgery when performed at levels distant from the DRF. Navigating instrumentation within 2 levels of the DRF likely minimizes the risk of navigation error. … (more)
- Is Part Of:
- Global spine journal. Volume 9:Issue 5(2019)
- Journal:
- Global spine journal
- Issue:
- Volume 9:Issue 5(2019)
- Issue Display:
- Volume 9, Issue 5 (2019)
- Year:
- 2019
- Volume:
- 9
- Issue:
- 5
- Issue Sort Value:
- 2019-0009-0005-0000
- Page Start:
- 512
- Page End:
- 520
- Publication Date:
- 2019-08
- Subjects:
- intraoperative navigation -- frameless stereotaxy -- image guidance -- registration -- navigation error
Spine -- Diseases -- Periodicals
Spine -- Diseases -- Treatment -- Periodicals
Spine -- Abnormalities -- Periodicals
Spine -- Surgery -- Periodicals
616.73 - Journal URLs:
- http://www.thieme.com/ ↗
- DOI:
- 10.1177/2192568218804556 ↗
- Languages:
- English
- ISSNs:
- 2192-5682
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 11107.xml