Optimization of an implantable magnetic marker for surgical localization of breast cancer. (7th September 2018)
- Record Type:
- Journal Article
- Title:
- Optimization of an implantable magnetic marker for surgical localization of breast cancer. (7th September 2018)
- Main Title:
- Optimization of an implantable magnetic marker for surgical localization of breast cancer
- Authors:
- Schermers, B
ten Haken, B
Muller, S H
van der Hage, J A
Ruers, T J M - Abstract:
- Abstract: For small, early-stage or otherwise non-palpable breast tumors, surgeons rely on localization technologies to accurately find and remove the tumor tissue during breast conserving surgery. However, current widely accepted localization technologies either use painful and logistically challenging guidewires, or complex radioactive iodine sources. We have developed an implantable magnetic marker, intended to mark the location of a breast tumor, that can be detected during surgery using a clinical handheld magnetic susceptometry system. Here, we report on the development and optimization of this magnetic marker, focusing on the material, shape and various material assemblies. It was found that the effects of magnetic shape anisotropy may decrease localization precision. This can be circumvented by combining multiple isotropic magnetic elements separated from one another. A final optimized prototype was constructed and compared to a commercially available magnetic marker. Finally, the technology was tested in an ex vivo surgical setting on tissue to assess radiological visibility and surgical feasibility. The marker was successfully detected and removed in all ex vivo sessions, and the technology was found feasible.
- Is Part Of:
- Biomedical physics & engineering express. Volume 4:Number 6(2018)
- Journal:
- Biomedical physics & engineering express
- Issue:
- Volume 4:Number 6(2018)
- Issue Display:
- Volume 4, Issue 6 (2018)
- Year:
- 2018
- Volume:
- 4
- Issue:
- 6
- Issue Sort Value:
- 2018-0004-0006-0000
- Page Start:
- Page End:
- Publication Date:
- 2018-09-07
- Subjects:
- surgical localization -- magnetic localization -- magnetic anisotropy -- magnetic shape anisotropy -- breast cancer -- cancer surgery -- oncological surgery
Medical physics -- Periodicals
Biophysics -- Periodicals
Biomedical engineering -- Periodicals
Medical sciences -- Periodicals
610.153 - Journal URLs:
- http://iopscience.iop.org/2057-1976/ ↗
http://www.iop.org/ ↗ - DOI:
- 10.1088/2057-1976/aadb53 ↗
- Languages:
- English
- ISSNs:
- 2057-1976
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 11102.xml