Monte Carlo-based SPECT reconstruction within the SIMIND framework. (12th December 2018)
- Record Type:
- Journal Article
- Title:
- Monte Carlo-based SPECT reconstruction within the SIMIND framework. (12th December 2018)
- Main Title:
- Monte Carlo-based SPECT reconstruction within the SIMIND framework
- Authors:
- Gustafsson, Johan
Brolin, Gustav
Ljungberg, Michael - Abstract:
- Abstract: This paper presents the development and validation of a Monte Carlo-based singe photon emission computed tomography reconstruction program for parallel-hole collimation contained within the SIMIND Monte Carlo framework. The Monte Carlo code is used as an accurate forward-projector and is combined with a simplified back-projector to perform iterative tomographic reconstruction using the Maximum Likelihood Expectation Maximization and Ordered Subsets Expectation Maximization algorithms, together forming a program called SIMREC. The Monte Carlo simulation transforms the estimated source distribution directly from activity to counts in its projections. Hence, the reconstructed image is expressed in activity without reference to an external calibration. The program is tested using phantom measurements of spheres filled with 99m Tc, 177 Lu and 131 I placed in air and centrally and peripherally in a water-filled elliptical phantom. The feasibility of applying the reconstruction to patients is also demonstrated for a range of radiopharmaceuticals. The deviation in total activity in the spheres ranged between −4.1% and 6.2% compared with the activity determined when preparing the phantom. The SIMREC program was found to be accurate with respect to activity estimation and to reconstruct visually acceptable images within a few hours when applied to patient examples.
- Is Part Of:
- Physics in medicine & biology. Volume 63:Number 24(2018:Dec.)
- Journal:
- Physics in medicine & biology
- Issue:
- Volume 63:Number 24(2018:Dec.)
- Issue Display:
- Volume 63, Issue 24 (2018)
- Year:
- 2018
- Volume:
- 63
- Issue:
- 24
- Issue Sort Value:
- 2018-0063-0024-0000
- Page Start:
- Page End:
- Publication Date:
- 2018-12-12
- Subjects:
- SPECT -- Monte Carlo simulation -- Monte Carlo-based reconstruction -- quantitative SPECT
Biophysics -- Periodicals
Medical physics -- Periodicals
610.153 - Journal URLs:
- http://ioppublishing.org/ ↗
http://iopscience.iop.org/0031-9155 ↗ - DOI:
- 10.1088/1361-6560/aaf0f1 ↗
- Languages:
- English
- ISSNs:
- 0031-9155
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 11091.xml