Direct and indirect entrance surface dose measurement in X-ray diagnostics using nanoDot OSL dosimeters. (June 2019)
- Record Type:
- Journal Article
- Title:
- Direct and indirect entrance surface dose measurement in X-ray diagnostics using nanoDot OSL dosimeters. (June 2019)
- Main Title:
- Direct and indirect entrance surface dose measurement in X-ray diagnostics using nanoDot OSL dosimeters
- Authors:
- Musa, Yahaya
Hashim, Suhairul
Khalis Abdul Karim, Muhammad - Abstract:
- Abstract: This study investigated the suitability of nanoDot optically stimulated luminescence (OSL) dosimeters for entrance surface dose (ESD) measurements in common X-ray diagnostics. OSL dosimetry system supplied by Landauer Inc, (Glenwood, IL, USA) consisting of nanoDot OSL dosimeters and microStar reader were used. ESDs were measured with nanoDots placed on whole-body anthropomorphic phantom (for direct measurements) and in free air without backscatter material (for indirect measurements) in accordance with the IAEA Technical Report Series No. 457. Measurements were carried out in five X-ray projections including AP abdomen, AP chest, PA chest, AP thoracic spine and AP skull. The mean ESD for each projection was evaluated from multiple measurements with three different field sizes. Mean ESDs (direct measurements) for AP abdomen, AP chest, PA chest, AP thoracic spine and AP skull were 4.3 mGy, 0.7 mGy, 0.3 mGy, 6.8 mGy and 2.8 mGy respectively. While in the case of indirect measurements, the corresponding mean ESD values were 4.8 mGy, 0.7 mGy, 0.2 mGy, 7.2 mGy and 3.7 mGy respectively. The results were found to be within IAEA and European Commission (EC) diagnostic reference levels (DRLs) range. The percentage difference of the ESDs measured by direct and indirect methods in all projections were between 0 to 40%. It was affirmed that nanoDot OSL dosimeter is a good candidate for ESD measurement in common radiographic examinations, while taking into account the effect ofAbstract: This study investigated the suitability of nanoDot optically stimulated luminescence (OSL) dosimeters for entrance surface dose (ESD) measurements in common X-ray diagnostics. OSL dosimetry system supplied by Landauer Inc, (Glenwood, IL, USA) consisting of nanoDot OSL dosimeters and microStar reader were used. ESDs were measured with nanoDots placed on whole-body anthropomorphic phantom (for direct measurements) and in free air without backscatter material (for indirect measurements) in accordance with the IAEA Technical Report Series No. 457. Measurements were carried out in five X-ray projections including AP abdomen, AP chest, PA chest, AP thoracic spine and AP skull. The mean ESD for each projection was evaluated from multiple measurements with three different field sizes. Mean ESDs (direct measurements) for AP abdomen, AP chest, PA chest, AP thoracic spine and AP skull were 4.3 mGy, 0.7 mGy, 0.3 mGy, 6.8 mGy and 2.8 mGy respectively. While in the case of indirect measurements, the corresponding mean ESD values were 4.8 mGy, 0.7 mGy, 0.2 mGy, 7.2 mGy and 3.7 mGy respectively. The results were found to be within IAEA and European Commission (EC) diagnostic reference levels (DRLs) range. The percentage difference of the ESDs measured by direct and indirect methods in all projections were between 0 to 40%. It was affirmed that nanoDot OSL dosimeter is a good candidate for ESD measurement in common radiographic examinations, while taking into account the effect of energy dependence. … (more)
- Is Part Of:
- Journal of physics. Volume 1248(2019)
- Journal:
- Journal of physics
- Issue:
- Volume 1248(2019)
- Issue Display:
- Volume 1248, Issue 1 (2019)
- Year:
- 2019
- Volume:
- 1248
- Issue:
- 1
- Issue Sort Value:
- 2019-1248-0001-0000
- Page Start:
- Page End:
- Publication Date:
- 2019-06
- Subjects:
- Physics -- Congresses
530.5 - Journal URLs:
- http://www.iop.org/EJ/journal/1742-6596 ↗
http://ioppublishing.org/ ↗ - DOI:
- 10.1088/1742-6596/1248/1/012014 ↗
- Languages:
- English
- ISSNs:
- 1742-6588
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 5036.223000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 11096.xml