The effect of different CT scanners, scan parameters and scanning setup on Hounsfield units and calibrated bone density: a phantom study. (7th August 2018)
- Record Type:
- Journal Article
- Title:
- The effect of different CT scanners, scan parameters and scanning setup on Hounsfield units and calibrated bone density: a phantom study. (7th August 2018)
- Main Title:
- The effect of different CT scanners, scan parameters and scanning setup on Hounsfield units and calibrated bone density: a phantom study
- Authors:
- Free, Jeffrey
Eggermont, Florieke
Derikx, Loes
van Leeuwen, Ruud
van der Linden, Yvette
Jansen, Wim
Raaijmakers, Esther
Tanck, Esther
Kaatee, Robert - Abstract:
- Abstract: This study investigated the effect of different CT scanners, reconstruction protocols, scan positions, and air gaps on HU and/or calibrated calcium hydroxyapatite concentrations (CaHA). Phantoms containing bone-like materials were scanned on four different CT scanners. The effect of slice thickness, field of view (FOV) and reconstruction kernel on HU was investigated. Using clinical scan and reconstruction protocols and a calibration phantom, HU and CaHA were determined for different positions, and air gaps between phantom and calibration phantom. Changing reconstruction kernel considerably affected the HU (range −31 HU to 64 HU), whereas slice thickness and FOV did not. Inter-scanner differences in HU were <88 HU and decreased to <47 CaHA after calibration. Different positions of the phantom resulted in differences (on average up to −26 HU and −24 CaHA). An air gap of 2.5 cm atop the calibration phantom resulted in errors up to −41 CaHA. If absolute HU and calibrated CaHA are needed, different reconstruction kernels and changes in position within the FOV as well as air gaps should be avoided.
- Is Part Of:
- Biomedical physics & engineering express. Volume 4:Number 5(2018)
- Journal:
- Biomedical physics & engineering express
- Issue:
- Volume 4:Number 5(2018)
- Issue Display:
- Volume 4, Issue 5 (2018)
- Year:
- 2018
- Volume:
- 4
- Issue:
- 5
- Issue Sort Value:
- 2018-0004-0005-0000
- Page Start:
- Page End:
- Publication Date:
- 2018-08-07
- Subjects:
- CT scanner -- Hounsfield units -- calcium hydroxyapatite -- CT protocols -- inter-scanner differences
Medical physics -- Periodicals
Biophysics -- Periodicals
Biomedical engineering -- Periodicals
Medical sciences -- Periodicals
610.153 - Journal URLs:
- http://iopscience.iop.org/2057-1976/ ↗
http://www.iop.org/ ↗ - DOI:
- 10.1088/2057-1976/aad66a ↗