Measurement of the front-end dead-time of the LHCb muon detector and evaluation of its contribution to the muon detection inefficiency. (6th April 2016)
- Record Type:
- Journal Article
- Title:
- Measurement of the front-end dead-time of the LHCb muon detector and evaluation of its contribution to the muon detection inefficiency. (6th April 2016)
- Main Title:
- Measurement of the front-end dead-time of the LHCb muon detector and evaluation of its contribution to the muon detection inefficiency
- Authors:
- Anderlini, L.
Anelli, M.
Archilli, F.
Auriemma, G.
Baldini, W.
Bencivenni, G.
Bizzeti, A.
Bocci, V.
Bondar, N.
Bonivento, W.
Bochin, B.
Bozzi, C.
Brundu, D.
Cadeddu, S.
Campana, P.
Carboni, G.
Cardini, A.
Carletti, M.
Casu, L.
Chubykin, A.
Ciambrone, P.
Dané, E.
Simone, P. De
Falabella, A.
Felici, G.
Fiore, M.
Fontana, M.
Fresch, P.
Furfaro, E.
Graziani, G.
Kashchuk, A.
Kotriakhova, S.
Lai, A.
Lanfranchi, G.
Loi, A.
Maev, O.
Manca, G.
Martellotti, G.
Neustroev, P.
Oldeman, R.G.C.
Palutan, M.
Passaleva, G.
Penso, G.
Pinci, D.
Polycarpo, E.
Saitta, B.
Santacesaria, R.
Santimaria, M.
Santovetti, E.
Saputi, A.
Sarti, A.
Satriano, C.
Satta, A.
Schmidt, B.
Schneider, T.
Sciascia, B.
Sciubba, A.
Siddi, B.G.
Tellarini, G.
Vacca, C.
Vazquez-Gomez, R.
Vecchi, S.
Veltri, M.
Vorobyev, A.
… (more) - Abstract:
- Abstract: A method is described which allows to deduce the dead-time of the front-end electronics of the LHCb muon detector from a series of measurements performed at different luminosities at a bunch-crossing rate of 20 MHz. The measured values of the dead-time range from ∼ 70 ns to ∼ 100 ns. These results allow to estimate the performance of the muon detector at the future bunch-crossing rate of 40 MHz and at higher luminosity.
- Is Part Of:
- Journal of instrumentation. Volume 11:Number 4(2016:Apr.)
- Journal:
- Journal of instrumentation
- Issue:
- Volume 11:Number 4(2016:Apr.)
- Issue Display:
- Volume 11, Issue 4 (2016)
- Year:
- 2016
- Volume:
- 11
- Issue:
- 4
- Issue Sort Value:
- 2016-0011-0004-0000
- Page Start:
- P04010
- Page End:
- P04010
- Publication Date:
- 2016-04-06
- Subjects:
- Front-end electronics for detector readout -- Gaseous detectors -- Muon spectrometers -- Particle tracking detectors (Gaseous detectors) -- Wire chambers (MWPC, Thin-gap chambers, drift chambers, drift tubes, proportional chambers etc)
Scientific apparatus and instruments -- Periodicals
502.84 - Journal URLs:
- http://iopscience.iop.org/1748-0221 ↗
http://ioppublishing.org/ ↗ - DOI:
- 10.1088/1748-0221/11/04/P04010 ↗
- Languages:
- English
- ISSNs:
- 1748-0221
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 11068.xml