Dynamic susceptibility contrast parametric imaging using accelerated dual‐contrast echo planar imaging with keyhole. Issue 2 (8th January 2019)
- Record Type:
- Journal Article
- Title:
- Dynamic susceptibility contrast parametric imaging using accelerated dual‐contrast echo planar imaging with keyhole. Issue 2 (8th January 2019)
- Main Title:
- Dynamic susceptibility contrast parametric imaging using accelerated dual‐contrast echo planar imaging with keyhole
- Authors:
- Caldeira, Liliana L.
Yun, Seong Dae
da Silva, Nuno A.
Filss, Christian
Shah, N. Jon - Abstract:
- Abstract : Background: Echo planar imaging (EPI) is one of the methods of choice in dynamic susceptibility contrast MRI (DSC‐MRI) because it provides a sufficient temporal resolution. However, the relatively long readout duration of EPI often imposes limitations on increased spatial coverage or the use of multiple contrasts. Purpose: To develop a DSC‐MRI method using EPIK (EPI with keyhole) to provide dual‐contrast (TE1 and TE2 ) information with a higher spatial coverage than EPI. To compare results from the community‐standard EPI method and the proposed EPIK method. Study Type: Prospective. Subjects: One healthy subject and 17 brain tumor patients. Field Strength/Sequence: 3 T/accelerated EPI and dual‐contrast EPIK sequences. Assessment: After an initial evaluation using healthy in vivo images, the use of the proposed method for DSC‐MRI was verified with brain tumor patients. The parametric images (eg, CBF and CBV) and arterial input function (AIF), obtained from both the EPI and EPIK, were compared. Statistical Tests: The ratio of AIF peak height of the proposed method to that of EPI was computed. The ratio computation was also performed for the time‐to‐peak (TTP) in the AIF curves. From the obtained CBF and CBV maps, the tumor‐to‐brain (TBR) ratio was also calculated for each imaging method and the results were compared. Results: For the same temporal resolution (1.5 sec), EPIK yielded dual‐contrast (TEs of 13/33 msec) with an increased spatial coverage (24 slices) andAbstract : Background: Echo planar imaging (EPI) is one of the methods of choice in dynamic susceptibility contrast MRI (DSC‐MRI) because it provides a sufficient temporal resolution. However, the relatively long readout duration of EPI often imposes limitations on increased spatial coverage or the use of multiple contrasts. Purpose: To develop a DSC‐MRI method using EPIK (EPI with keyhole) to provide dual‐contrast (TE1 and TE2 ) information with a higher spatial coverage than EPI. To compare results from the community‐standard EPI method and the proposed EPIK method. Study Type: Prospective. Subjects: One healthy subject and 17 brain tumor patients. Field Strength/Sequence: 3 T/accelerated EPI and dual‐contrast EPIK sequences. Assessment: After an initial evaluation using healthy in vivo images, the use of the proposed method for DSC‐MRI was verified with brain tumor patients. The parametric images (eg, CBF and CBV) and arterial input function (AIF), obtained from both the EPI and EPIK, were compared. Statistical Tests: The ratio of AIF peak height of the proposed method to that of EPI was computed. The ratio computation was also performed for the time‐to‐peak (TTP) in the AIF curves. From the obtained CBF and CBV maps, the tumor‐to‐brain (TBR) ratio was also calculated for each imaging method and the results were compared. Results: For the same temporal resolution (1.5 sec), EPIK yielded dual‐contrast (TEs of 13/33 msec) with an increased spatial coverage (24 slices) and less geometric distortions than EPI; EPI provided single contrast (TE of 32 msec) with 20 slices. The obtained parametric values (eg, AIF peak, TTP, and TBR) had similar characteristics between EPI and the proposed method. Data Conclusion: The dual‐contrast data produced by EPIK in DSC‐MRI allowed T1 ‐corrected parametric images without the need of second contrast injection and an enhanced estimation of the AIF. Level of Evidence : 1 Technical Efficacy : Stage 1 J. Magn. Reson. Imaging 2019;50:628–640. … (more)
- Is Part Of:
- Journal of magnetic resonance imaging. Volume 50:Issue 2(2019)
- Journal:
- Journal of magnetic resonance imaging
- Issue:
- Volume 50:Issue 2(2019)
- Issue Display:
- Volume 50, Issue 2 (2019)
- Year:
- 2019
- Volume:
- 50
- Issue:
- 2
- Issue Sort Value:
- 2019-0050-0002-0000
- Page Start:
- 628
- Page End:
- 640
- Publication Date:
- 2019-01-08
- Subjects:
- AIF -- dual‐contrast -- EPIK -- image distortion -- increased spatial coverage -- leakage correction
Magnetic resonance imaging -- Periodicals
616 - Journal URLs:
- http://onlinelibrary.wiley.com/journal/10.1002/(ISSN)1522-2586 ↗
http://onlinelibrary.wiley.com/ ↗ - DOI:
- 10.1002/jmri.26639 ↗
- Languages:
- English
- ISSNs:
- 1053-1807
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 5010.791000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 11039.xml