Dose distribution of secondary radiation in a water phantom for a proton pencil beam—EURADOS WG9 intercomparison exercise. (19th April 2018)
- Record Type:
- Journal Article
- Title:
- Dose distribution of secondary radiation in a water phantom for a proton pencil beam—EURADOS WG9 intercomparison exercise. (19th April 2018)
- Main Title:
- Dose distribution of secondary radiation in a water phantom for a proton pencil beam—EURADOS WG9 intercomparison exercise
- Authors:
- Stolarczyk, L
Trinkl, S
Romero-Expósito, M
Mojżeszek, N
Ambrozova, I
Domingo, C
Davídková, M
Farah, J
Kłodowska, M
Knežević, Ž
Liszka, M
Majer, M
Miljanić, S
Ploc, O
Schwarz, M
Harrison, R M
Olko, P - Abstract:
- Abstract: Systematic 3D mapping of out-of-field doses induced by a therapeutic proton pencil scanning beam in a 300 × 300 × 600 mm 3 water phantom was performed using a set of thermoluminescence detectors (TLDs): MTS-7 ( 7 LiF:Mg, Ti), MTS-6 ( 6 LiF:Mg, Ti), MTS-N ( nat LiF:Mg, Ti) and TLD-700 ( 7 LiF:Mg, Ti), radiophotoluminescent (RPL) detectors GD-352M and GD-302M, and polyallyldiglycol carbonate (PADC)-based (C12 H18 O7 ) track-etched detectors. Neutron and gamma-ray doses, as well as linear energy transfer distributions, were experimentally determined at 200 points within the phantom. In parallel, the Geant4 Monte Carlo code was applied to calculate neutron and gamma radiation spectra at the position of each detector. For the cubic proton target volume of 100 × 100 × 100 mm 3 (spread out Bragg peak with a modulation of 100 mm) the scattered photon doses along the main axis of the phantom perpendicular to the primary beam were approximately 0.5 mGy Gy −1 at a distance of 100 mm and 0.02 mGy Gy −1 at 300 mm from the center of the target. For the neutrons, the corresponding values of dose equivalent were found to be ~0.7 and ~0.06 mSv Gy −1, respectively. The measured neutron doses were comparable with the out-of-field neutron doses from a similar experiment with 20 MV x-rays, whereas photon doses for the scanning proton beam were up to three orders of magnitude lower.
- Is Part Of:
- Physics in medicine & biology. Volume 63:Number 8(2018:Apr.)
- Journal:
- Physics in medicine & biology
- Issue:
- Volume 63:Number 8(2018:Apr.)
- Issue Display:
- Volume 63, Issue 8 (2018)
- Year:
- 2018
- Volume:
- 63
- Issue:
- 8
- Issue Sort Value:
- 2018-0063-0008-0000
- Page Start:
- Page End:
- Publication Date:
- 2018-04-19
- Subjects:
- passive detectors -- neutron dosimetry -- gamma radiation dosimetry -- water phantom measurements -- secondary radiation measurements -- pencil beam scanning proton radiotherapy
Biophysics -- Periodicals
Medical physics -- Periodicals
610.153 - Journal URLs:
- http://ioppublishing.org/ ↗
http://iopscience.iop.org/0031-9155 ↗ - DOI:
- 10.1088/1361-6560/aab469 ↗
- Languages:
- English
- ISSNs:
- 0031-9155
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 10955.xml