A mesh nebulizer is more effective than jet nebulizer to nebulize bronchodilators during non-invasive ventilation of subjects with COPD: A randomized controlled trial with radiolabeled aerosols. (July 2019)
- Record Type:
- Journal Article
- Title:
- A mesh nebulizer is more effective than jet nebulizer to nebulize bronchodilators during non-invasive ventilation of subjects with COPD: A randomized controlled trial with radiolabeled aerosols. (July 2019)
- Main Title:
- A mesh nebulizer is more effective than jet nebulizer to nebulize bronchodilators during non-invasive ventilation of subjects with COPD: A randomized controlled trial with radiolabeled aerosols
- Authors:
- Galindo-Filho, Valdecir Castor
Alcoforado, Luciana
Rattes, Catarina
Paiva, Dulciane Nunes
Brandão, Simone Cristina Soares
Fink, James B.
Dornelas de Andrade, Armèle - Abstract:
- Abstract: Background: Beneficial effects from non-invasive ventilation (NIV) in acute COPD are well-established, but the impact of nebulization during NIV has not been well described. Aim: To compare pulmonary deposition and distribution across regions of interest with administration of radiolabeled aerosols generated by vibrating mesh nebulizers (VMN) and jet nebulizer (JN) during NIV. Methods: A crossover single dose study involving 9 stable subjects with moderate to severe COPD randomly allocated to receive aerosol administration by the VMN Aerogen and the MistyNeb jet nebulizer operating with oxygen at 8 lpm during NIV. Radiolabeled bronchodilators (fill volume of 3 mL: 0.5 mL salbutamol 2.5 mg + 0.125 mL ipratropium 0.25 mg and physiologic saline up to 3 mL) were delivered until sputtering during NIV (pressures of 12 cmH2O and 5 cmH2O - inspiratory and expiratory, respectively) using an oro-nasal facemask. Radioactivity counts were performed using a gamma camera and regions of interest (ROIs) were delimited. Aerosol mass balance based on counts from the lungs, upper airways, stomach, nebulizer, circuit, inspiratory and expiratory filters, and mask were determined and expressed as a percentage of the total. Results: Both inhaled and lung doses were greater with VMN (22.78 ± 3.38% and 12.05 ± 2.96%, respectively) than JN (12.51 ± 6.31% and 3.14 ± 1.71%; p = 0.008). Residual drug volume was lower in VMN than in JN (3.08 ± 1.3% versus 46.44 ± 5.83%, p = 0.001). PeripheralAbstract: Background: Beneficial effects from non-invasive ventilation (NIV) in acute COPD are well-established, but the impact of nebulization during NIV has not been well described. Aim: To compare pulmonary deposition and distribution across regions of interest with administration of radiolabeled aerosols generated by vibrating mesh nebulizers (VMN) and jet nebulizer (JN) during NIV. Methods: A crossover single dose study involving 9 stable subjects with moderate to severe COPD randomly allocated to receive aerosol administration by the VMN Aerogen and the MistyNeb jet nebulizer operating with oxygen at 8 lpm during NIV. Radiolabeled bronchodilators (fill volume of 3 mL: 0.5 mL salbutamol 2.5 mg + 0.125 mL ipratropium 0.25 mg and physiologic saline up to 3 mL) were delivered until sputtering during NIV (pressures of 12 cmH2O and 5 cmH2O - inspiratory and expiratory, respectively) using an oro-nasal facemask. Radioactivity counts were performed using a gamma camera and regions of interest (ROIs) were delimited. Aerosol mass balance based on counts from the lungs, upper airways, stomach, nebulizer, circuit, inspiratory and expiratory filters, and mask were determined and expressed as a percentage of the total. Results: Both inhaled and lung doses were greater with VMN (22.78 ± 3.38% and 12.05 ± 2.96%, respectively) than JN (12.51 ± 6.31% and 3.14 ± 1.71%; p = 0.008). Residual drug volume was lower in VMN than in JN (3.08 ± 1.3% versus 46.44 ± 5.83%, p = 0.001). Peripheral deposition of radioaerosol was significantly lower with JN than VMN. Conclusions: VMN deposited > 3 fold more radioaerosol into the lungs of moderate to severe COPD patients than JN during NIV. Highlights: The mesh nebulizer (VMN) provides similar or greater lung deposition than standard jet nebulizers (JN). JN delivers < 25% of lung deposition than VMN during NIV. In spontaneously breathing patients, suggesting that standard label doses could be applied with Mesh during NIV. … (more)
- Is Part Of:
- Respiratory medicine. Volume 153(2019)
- Journal:
- Respiratory medicine
- Issue:
- Volume 153(2019)
- Issue Display:
- Volume 153, Issue 2019 (2019)
- Year:
- 2019
- Volume:
- 153
- Issue:
- 2019
- Issue Sort Value:
- 2019-0153-2019-0000
- Page Start:
- 60
- Page End:
- 67
- Publication Date:
- 2019-07
- Subjects:
- Vibrating mesh nebulizers -- Jet nebulizer -- Noninvasive ventilation -- Pulmonary scintigraphy -- Radioaerosol -- COPD
Chest -- Diseases -- Periodicals
Chest -- Diseases -- Great Britain -- Periodicals
Respiratory organs -- Diseases -- Periodicals
Respiratory Tract Diseases -- Periodicals
Appareil respiratoire -- Maladies -- Périodiques
Thorax -- Maladies -- Périodiques
Appareil respiratoire -- Maladies -- Traitement -- Périodiques
Electronic journals
616.2 - Journal URLs:
- http://www.sciencedirect.com/science/journal/09546111 ↗
http://www.clinicalkey.com/dura/browse/journalIssue/09546111 ↗
http://www.clinicalkey.com.au/dura/browse/journalIssue/09546111 ↗
http://www.elsevier.com/journals ↗ - DOI:
- 10.1016/j.rmed.2019.05.016 ↗
- Languages:
- English
- ISSNs:
- 0954-6111
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 7777.661900
British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 10950.xml