Biomechanical performance of cranial implants with different thicknesses and material properties: A finite element study. (June 2019)
- Record Type:
- Journal Article
- Title:
- Biomechanical performance of cranial implants with different thicknesses and material properties: A finite element study. (June 2019)
- Main Title:
- Biomechanical performance of cranial implants with different thicknesses and material properties: A finite element study
- Authors:
- Marcián, Petr
Narra, Nathaniel
Borák, Libor
Chamrad, Jakub
Wolff, Jan - Abstract:
- Abstract: This study investigated the effect of implant thickness and material on deformation and stress distribution within different components of cranial implant assemblies. Using the finite element method, two cranial implants, differing in size and shape, and thicknesses (1, 2, 3 and 4 mm, respectively), were simulated under three loading scenarios. The implant assembly model included the detailed geometries of the mini-plates and micro-screws and was simulated using a sub-modeling approach. Statistical assessments based on the Design of Experiment methodology and on multiple regression analysis revealed that peak stresses in the components are influenced primarily by implant thickness, while the effect of implant material is secondary. On the contrary, the implant deflection is influenced predominantly by implant material followed by implant thickness. The highest values of deformation under a 50 N load were observed in the thinnest (1 mm) Polymethyl Methacrylate implant (Small defect: 0.296 mm; Large defect: 0.390 mm). The thinnest Polymethyl Methacrylate and Polyether Ether Ketone implants also generated stresses in the implants that can potentially breach the materials' yield limit. In terms of stress distribution, the change of implant thickness had a more significant impact on the implant performance than the change of Young's modulus of the implant material. The results indicated that the stresses are concentrated in the locations of fixation; therefore, theAbstract: This study investigated the effect of implant thickness and material on deformation and stress distribution within different components of cranial implant assemblies. Using the finite element method, two cranial implants, differing in size and shape, and thicknesses (1, 2, 3 and 4 mm, respectively), were simulated under three loading scenarios. The implant assembly model included the detailed geometries of the mini-plates and micro-screws and was simulated using a sub-modeling approach. Statistical assessments based on the Design of Experiment methodology and on multiple regression analysis revealed that peak stresses in the components are influenced primarily by implant thickness, while the effect of implant material is secondary. On the contrary, the implant deflection is influenced predominantly by implant material followed by implant thickness. The highest values of deformation under a 50 N load were observed in the thinnest (1 mm) Polymethyl Methacrylate implant (Small defect: 0.296 mm; Large defect: 0.390 mm). The thinnest Polymethyl Methacrylate and Polyether Ether Ketone implants also generated stresses in the implants that can potentially breach the materials' yield limit. In terms of stress distribution, the change of implant thickness had a more significant impact on the implant performance than the change of Young's modulus of the implant material. The results indicated that the stresses are concentrated in the locations of fixation; therefore, the detailed models of mini-plates and micro-screws implemented in the finite element simulation provided a better insight into the mechanical performance of the implant-skull system. Graphical abstract: Image 1 Highlights: Mechanical stresses in cranial implants. Micro-screw geometry has a significant effect on cranial implant stresses. Implant deflection is predominantly influenced by the cranial implant material. Stress within cranial implant is predominantly influenced by the implant thickness. Implant thickness has a higher effect on stresses than material of cranial implant. … (more)
- Is Part Of:
- Computers in biology and medicine. Volume 109(2019)
- Journal:
- Computers in biology and medicine
- Issue:
- Volume 109(2019)
- Issue Display:
- Volume 109, Issue 2019 (2019)
- Year:
- 2019
- Volume:
- 109
- Issue:
- 2019
- Issue Sort Value:
- 2019-0109-2019-0000
- Page Start:
- 43
- Page End:
- 52
- Publication Date:
- 2019-06
- Subjects:
- Cranioplasty -- Skull implant -- 3D printing -- Finite element method -- Mechanical properties
Medicine -- Data processing -- Periodicals
Biology -- Data processing -- Periodicals
610.285 - Journal URLs:
- http://www.sciencedirect.com/science/journal/00104825/ ↗
http://www.elsevier.com/journals ↗ - DOI:
- 10.1016/j.compbiomed.2019.04.016 ↗
- Languages:
- English
- ISSNs:
- 0010-4825
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 3394.880000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 10932.xml