Comparison of treatment planning approaches for spatially fractionated irradiation of deep tumors. Issue 6 (21st May 2019)
- Record Type:
- Journal Article
- Title:
- Comparison of treatment planning approaches for spatially fractionated irradiation of deep tumors. Issue 6 (21st May 2019)
- Main Title:
- Comparison of treatment planning approaches for spatially fractionated irradiation of deep tumors
- Authors:
- Sheikh, Khadija
Hrinivich, William T.
Bell, Leslie A.
Moore, Joseph A.
Laub, Wolfram
Viswanathan, Akila N.
Yan, Yulong
McNutt, Todd R.
Meyer, Jeffrey - Abstract:
- Abstract: Purpose: The purpose of this work was to compare the dosimetry and delivery times of 3D‐conformal (3DCRT)‐, volumetric modulated arc therapy (VMAT)‐, and tomotherapy‐based approaches for spatially fractionated radiation therapy for deep tumor targets. Methods: Two virtual GRID phantoms were created consisting of 7 "target" cylinders (1‐cm diameter) aligned longitudinally along the tumor in a honey‐comb pattern, mimicking a conventional GRID block, with 2‐cm center‐to‐center spacing (GRID2 cm ) and 3‐cm center‐to‐center spacing (GRID3 cm ), all contained within a larger cylinder (8 and 10 cm in diameter for the GRID2 cm and GRID3 cm, respectively). In a single patient, a GRID3 cm structure was created within the gross tumor volume (GTV). Tomotherapy, VMAT (6 MV + 6 MV‐flattening‐filter‐free) and multi‐leaf collimator segment 3DCRT (6 MV) plans were created using commercially available software. Two tomotherapy plans were created with field widths (TOMO2.5 cm ) 2.5 cm and (TOMO5 cm ) 5 cm. Prescriptions for all plans were set to deliver a mean dose of 15 Gy to the GRID targets in one fraction. The mean dose to the GRID target and the heterogeneity of the dose distribution (peak‐to‐valley and peak‐to‐edge dose ratios) inside the GRID target were obtained. The volume of normal tissue receiving 7.5 Gy was determined. Results: The peak‐to‐valley ratios for GRID2 cm /GRID3 cm /Patient were 2.1/2.3/2.8, 1.7/1.5/2.8, 1.7/1.9/2.4, and 1.8/2.0/2.8 for the 3DCRT, VMAT,Abstract: Purpose: The purpose of this work was to compare the dosimetry and delivery times of 3D‐conformal (3DCRT)‐, volumetric modulated arc therapy (VMAT)‐, and tomotherapy‐based approaches for spatially fractionated radiation therapy for deep tumor targets. Methods: Two virtual GRID phantoms were created consisting of 7 "target" cylinders (1‐cm diameter) aligned longitudinally along the tumor in a honey‐comb pattern, mimicking a conventional GRID block, with 2‐cm center‐to‐center spacing (GRID2 cm ) and 3‐cm center‐to‐center spacing (GRID3 cm ), all contained within a larger cylinder (8 and 10 cm in diameter for the GRID2 cm and GRID3 cm, respectively). In a single patient, a GRID3 cm structure was created within the gross tumor volume (GTV). Tomotherapy, VMAT (6 MV + 6 MV‐flattening‐filter‐free) and multi‐leaf collimator segment 3DCRT (6 MV) plans were created using commercially available software. Two tomotherapy plans were created with field widths (TOMO2.5 cm ) 2.5 cm and (TOMO5 cm ) 5 cm. Prescriptions for all plans were set to deliver a mean dose of 15 Gy to the GRID targets in one fraction. The mean dose to the GRID target and the heterogeneity of the dose distribution (peak‐to‐valley and peak‐to‐edge dose ratios) inside the GRID target were obtained. The volume of normal tissue receiving 7.5 Gy was determined. Results: The peak‐to‐valley ratios for GRID2 cm /GRID3 cm /Patient were 2.1/2.3/2.8, 1.7/1.5/2.8, 1.7/1.9/2.4, and 1.8/2.0/2.8 for the 3DCRT, VMAT, TOMO5 cm, and TOMO2.5 cm plans, respectively. The peak‐to‐edge ratios for GRID2 cm /GRID3 cm /Patient were 2.8/3.2/5.4, 2.1/1.8/5.4, 2.0/2.2/3.9, 2.1/2.7/5.2 and for the 3DCRT, VMAT, TOMO5 cm, and TOMO2.5 cm plans, respectively. The volume of normal tissue receiving 7.5 Gy was lowest in the TOMO2.5 cm plan (GRID2 cm /GRID3 cm /Patient = 54 cm 3 /19 cm 3 /10 cm 3 ). The VMAT plans had the lowest delivery times (GRID2 cm /GRID3 cm /Patient = 17 min/8 min/9 min). Conclusion: Our results present, for the first time, preliminary evidence comparing IMRT‐GRID approaches which result in high‐dose "islands" within a target, mimicking what is achieved with a conventional GRID block but without high‐dose "tail" regions outside of the target. These approaches differ modestly in their ability to achieve high peak‐to‐edge ratios and also differ in delivery times. … (more)
- Is Part Of:
- Journal of applied clinical medical physics. Volume 20:Issue 6(2019)
- Journal:
- Journal of applied clinical medical physics
- Issue:
- Volume 20:Issue 6(2019)
- Issue Display:
- Volume 20, Issue 6 (2019)
- Year:
- 2019
- Volume:
- 20
- Issue:
- 6
- Issue Sort Value:
- 2019-0020-0006-0000
- Page Start:
- 125
- Page End:
- 133
- Publication Date:
- 2019-05-21
- Subjects:
- deep‐seated tumors -- grid therapy -- spatially fractionated radiotherapy
Medical physics -- Periodicals
Clinical medicine -- Periodicals
Health Physics
Clinical Medicine
Electronic journals
Periodicals
Periodicals
Fulltext
Internet Resources
610.153 - Journal URLs:
- http://aapm.onlinelibrary.wiley.com/hub/journal/10.1002/(ISSN)1526-9914/ ↗
http://bibpurl.oclc.org/web/7294 ↗
http://www.jacmp.org/ ↗
http://onlinelibrary.wiley.com/ ↗ - DOI:
- 10.1002/acm2.12617 ↗
- Languages:
- English
- ISSNs:
- 1526-9914
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 10862.xml