A miniaturized, tethered, spectrally‐encoded confocal endomicroscopy capsule. Issue 5 (6th January 2019)
- Record Type:
- Journal Article
- Title:
- A miniaturized, tethered, spectrally‐encoded confocal endomicroscopy capsule. Issue 5 (6th January 2019)
- Main Title:
- A miniaturized, tethered, spectrally‐encoded confocal endomicroscopy capsule
- Authors:
- Kang, Dongkyun
Do, Dukho
Ryu, Jiheun
Grant, Catriona N.
Giddings, Sarah L.
Rosenberg, Mireille
Hesterberg, Paul E.
Yuan, Qian
Garber, John J.
Katz, Aubrey J.
Tearney, Guillermo J. - Other Names:
- Tearney Guillermo J. guestEditor.
Kang Dongkyun guestEditor.
Wang Thomas D. guestEditor. - Abstract:
- Abstract : Background and Objective: The tethered spectrally‐encoded confocal endomicroscopy (SECM) capsule is an imaging device that once swallowed by an unsedated patient can visualize cellular morphologic changes associated with gastrointestinal (GI) tract diseases in vivo . Recently, we demonstrated a tethered SECM capsule for counting esophageal eosinophils in patients with eosinophilic esophagitis (EoE) in vivo . Yet, the current tethered SECM capsule is far too long to be widely utilized for imaging pediatric patients, who constitute a major portion of the EoE patient population. In this paper, we present a new tethered SECM capsule that is 33% shorter, has an easier and repeatable fabrication process, and produces images with reduced speckle noise. Materials and Methods: The smaller SECM capsule utilized a miniature condenser to increase the fiber numerical aperture and reduce the capsule length. A custom 3D‐printed holder was developed to enable easy and repeatable device fabrication. A dual‐clad fiber (DCF) was used to reduce speckle noise. Results: The fabricated SECM capsule (length = 20 mm; diameter = 7 mm) had a similar size and shape to a pediatric dietary supplement pill. The new capsule achieved optical sectioning thickness of 13.2 μm with a small performance variation between devices of 1.7 μm. Confocal images of human esophagus obtained in vivo showed the capability of this new device to clearly resolve microstructural epithelial details with reducedAbstract : Background and Objective: The tethered spectrally‐encoded confocal endomicroscopy (SECM) capsule is an imaging device that once swallowed by an unsedated patient can visualize cellular morphologic changes associated with gastrointestinal (GI) tract diseases in vivo . Recently, we demonstrated a tethered SECM capsule for counting esophageal eosinophils in patients with eosinophilic esophagitis (EoE) in vivo . Yet, the current tethered SECM capsule is far too long to be widely utilized for imaging pediatric patients, who constitute a major portion of the EoE patient population. In this paper, we present a new tethered SECM capsule that is 33% shorter, has an easier and repeatable fabrication process, and produces images with reduced speckle noise. Materials and Methods: The smaller SECM capsule utilized a miniature condenser to increase the fiber numerical aperture and reduce the capsule length. A custom 3D‐printed holder was developed to enable easy and repeatable device fabrication. A dual‐clad fiber (DCF) was used to reduce speckle noise. Results: The fabricated SECM capsule (length = 20 mm; diameter = 7 mm) had a similar size and shape to a pediatric dietary supplement pill. The new capsule achieved optical sectioning thickness of 13.2 μm with a small performance variation between devices of 1.7 μm. Confocal images of human esophagus obtained in vivo showed the capability of this new device to clearly resolve microstructural epithelial details with reduced speckle noise. Conclusions: We expect that the smaller size and better image performance of this new SECM capsule will greatly facilitate the clinical adoption of this technology in pediatric patients and will enable more accurate assessment of EoE‐suspected tissues. Lasers Surg. Med. 51:452–458, 2019. © 2019 Wiley Periodicals, Inc. … (more)
- Is Part Of:
- Lasers in surgery and medicine. Volume 51:Issue 5(2019)
- Journal:
- Lasers in surgery and medicine
- Issue:
- Volume 51:Issue 5(2019)
- Issue Display:
- Volume 51, Issue 5 (2019)
- Year:
- 2019
- Volume:
- 51
- Issue:
- 5
- Issue Sort Value:
- 2019-0051-0005-0000
- Page Start:
- 452
- Page End:
- 458
- Publication Date:
- 2019-01-06
- Subjects:
- confocal endomicroscopy -- spectrally encoded confocal microscopy -- reflectance confocal microcopy -- tethered capsule endomicroscopy -- eosinophilic esophagitis -- esophageal imaging
Lasers in medicine -- Periodicals
Lasers in surgery -- Periodicals
617 - Journal URLs:
- http://onlinelibrary.wiley.com/ ↗
- DOI:
- 10.1002/lsm.23050 ↗
- Languages:
- English
- ISSNs:
- 0196-8092
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 5156.683000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 10862.xml